Feasibility of pseudocontinuous arterial spin labeling at 7 T with whole-brain coverage.

MAGMA

Department of Radiology, C.J. Gorter Center for High Field MRI, Leiden University Medical Center (LUMC), C3-Q, Albinusdreef 2, 2333 ZA Leiden, The Netherlands.

Published: April 2012

Object: We studied the feasibility of pseudocontinuous arterial spin labeling (pCASL) at 7 T.

Materials And Methods: Simulations were performed to find the optimal labeling parameters for pCASL, with particular attention to the maximum-allowed specific absorption rate (SAR). Subsequently, pCASL experiments (four volunteers) were performed to find the B1 efficiency at the labeling position with and without high-permittivity pads placed around the head, and to study the optimal labeling duration (four separate volunteers). Finally, feasibility of whole-brain pCASL imaging was tested.

Results: Simulations showed that a lower B1 efficiency should be compensated by a lower effective flip angle of the labeling, a moderately shorter labeling duration, and a longer repetition time. B1 efficiency in the internal carotid arteries just below the carotid siphon was approximately 55% and 35% with and without high-permittivity pads, respectively. In vivo experiments showed an optimal labeling duration of 1,500 ms, although longer labeling durations up to 2,500 ms resulted in similar signal-to-noise efficiency. Whole-brain pCASL imaging was demonstrated in a single volunteer.

Conclusion: Despite decreased B1 efficiency, sufficient labeling efficiency can be achieved for whole-brain pCASL at 7 T with high-permittivity pads. However, image quality is still limited compared with 3 T, probably due to imaging instabilities, and further research is needed to elucidate this.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3313026PMC
http://dx.doi.org/10.1007/s10334-011-0297-0DOI Listing

Publication Analysis

Top Keywords

optimal labeling
12
high-permittivity pads
12
labeling duration
12
whole-brain pcasl
12
labeling
10
feasibility pseudocontinuous
8
pseudocontinuous arterial
8
arterial spin
8
spin labeling
8
performed find
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!