Soft tissue balancing during total knee arthroplasty (TKA) has a direct affect on patello-femoral tracking and knee range of motion, which are necessary for a well functioning TKA postoperatively. We report on the use of an electronic pressure sensing instrument for soft tissue balancing of the knee before completion of all intraoperative bone cuts, as a way to improve patellar tracking. In a retrospective study of 99 consecutive TKAs, with intraoperative electronic instrument guided soft tissue balancing performed, a reduction in the incidence of lateral patellar retinacular release was found, as compared with the 100 consecutive TKAs prior to its use (5.5% v 12%, respectively). Electronic measurement of soft tissue balancing during TKA reduced the need for lateral patellar retinacular release.