Purpose: To evaluate the relative effectiveness and safety of LASIK using the Femto LDV (Ziemer Ophthalmic Systems AG) and IntraLase FS 60 (Abbott Medical Optics Inc) femtosecond lasers in patients with corneal opacity.

Methods: Patients with corneal opacity were retrospectively selected between March and July 2009. For this study, 205 eyes with 90-μm corneal flaps created using the Femto LDV (LDV group) and 200 eyes with corneal flaps created using the IntraLase FS 60 (Intra-Lase group) were selected. The flap thickness of the IntraLase group was determined by observation with slit-lamp microscopy. Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), and manifest refraction spherical equivalent (MRSE) were measured pre- and postoperatively and were statistically evaluated using the Student t test and Mann-Whitney U-test.

Results: Regardless of the levels of opacity, eyes in the LDV group experienced uneventful procedures with no complications. Eyes in the IntraLase group had corneal flaps of 100- to 130-μm thickness and uneventful procedures; however, gas breakthrough was observed in 27 eyes. Of all eyes, 117 eyes from the LDV group and 109 eyes from the IntraLase group were available for 3-month follow-up. Mean 3-month postoperative UDVA, CDVA, and MRSE for the LDV group were 20/12.5, 20/12.5, and 0.17±0.32 diopters (D), respectively, and for the IntraLase group were 20/12.5, 20/12.5, and 0.11±0.34 D, respectively. No statistically significant differenes were noted in UDVA, CDVA, or MRSE between groups (P>.05 for all).

Conclusions: Laser in situ keratomileusis with the Femto LDV created thin flaps regardless of level of opacity and induced no complications as compared to the IntraLase FS 60, where gas breakthrough was significantly more common.

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081597X-20111213-01DOI Listing

Publication Analysis

Top Keywords

femto ldv
16
ldv group
16
intralase group
16
patients corneal
12
corneal flaps
12
group
9
ldv
8
corneal opacity
8
eyes
8
flaps created
8

Similar Publications

Purpose: To present the outcomes and detail the surgical procedure employed in the initial 102 eyes treated with keratorefractive lenticule extraction (KLEx) using the FEMTO LDV Z8 platform (Ziemer Ophthalmic Systems).

Methods: This was a retrospective analysis of 102 eyes of 53 patients treated with KLEx at the American University of Beirut Medical Center. Visual, refractive, topographic, and aberrometric parameters were evaluated 1 week and 1, 3, 6, and 12 months postoperatively.

View Article and Find Full Text PDF

Background: Small incision lenticule extraction (SMILE) and femtosecond laser-assisted in situ keratomileusis (LASIK) are widely used surgical methods to correct myopia with comparable efficacy, predictability, and safety. We examined and compared the early changes of tear protein profiles after SMILE and FS-LASIK surgery in order to find possible differences in the initial corneal healing process.

Methods: SMILE operations for 26 eyes were made with Visumax femtosecond laser.

View Article and Find Full Text PDF

Introduction: Corneal lenticules can be utilized as an additive material for stromal keratophakia. However, following extraction, they must be reimplanted almost immediately or cryopreserved in lenticule banks. Electron-beam (E-beam) irradiated corneas permit room-temperature storage for up to 2 years, enabling keratophakia to be performed on demand.

View Article and Find Full Text PDF

Background: To compare the safety and efficacy of femtosecond laser-assisted cataract surgery (FLACS) performed with the low-energy FEMTO LDV Z8 (Ziemer Ophthalmic Systems AG, Port, Switzerland) laser compared with conventional phacoemulsification (CP) in Chinese patients.

Methods: This prospective, multicenter, interventional study included 126 patients who were randomized (1:1) to undergo either FLACS or CP followed by intraocular lens (IOL) implantation between January 2019 and April 2020. The primary endpoint included the comparison of the endothelial cell loss (ECL) between the two groups at 3 months.

View Article and Find Full Text PDF

Purpose: Currently, two major principles exist to create LASIK flaps: firstly, a strictly horizontal (2D) cut similar to the microkeratome-cut and secondly an angled cut with a "step-like" edge (3D). The strictly horizontal (2D) cut method can be performed using apparatus such as the low-energy FEMTO LDV Z8 laser and its predecessors which are specific to this type. Alternatively, the low-energy FEMTO LDV Z8 laser's 3D flap design creates an interlocking flap-interface surface which potentially contributes toward flap stability.

View Article and Find Full Text PDF