Purpose: To develop a new automatic exposure control (AEC) technique based on the contrast-to-noise ratio (CNR) and provide constant lesion detectability.

Methods: Lesion detectability is affected by factors such as image noise, lesion contrast, and lesion size. We performed ROC analysis to assess the relationship between the optimum CNR and the lesion diameter at various levels of lesion contrast. We then developed a CNR-based AEC algorithm based on lesion detectability. Using CNR- based AEC algorithm, we performed visual evaluation of low-contrast detectability by 5 radiologists on a low-contrast module of the Catphan phantom, a contrast-difference level of 1.0% (difference in the CT number = 10 HU), and objects 3.0-9.0 mm in diameter.

Results: On step-and-shoot scans the mean detection fraction with CNR-based AEC remained almost constant from 88 to 99 % regardless of the lesion size. We observed the same trend on helical scans, the mean detection fraction with CNR-based AEC exhibited a high score from 91 to 100%. Although CNR-based AEC maintains higher CNR for smaller size or lower contrast lesion, radiation dose on 3 mm lesion resulted in about 13 times larger than that of 9 mm lesion size. CTDI(vol) for the CNR-based AEC technique changed dramatically with the SD(Z) from 7.5 to 100.0 mGy for step-and-shoot scans and from 9.1 to 121.5 mGy for helical scans.

Conclusions: From the viewpoint of ROC analysis-based CNR for lesion detection, CNR-based AEC potentially provide image quality advantages for clinical implementation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejmp.2011.11.004DOI Listing

Publication Analysis

Top Keywords

cnr-based aec
24
lesion
12
lesion size
12
automatic exposure
8
exposure control
8
based contrast-to-noise
8
contrast-to-noise ratio
8
aec
8
aec technique
8
constant lesion
8

Similar Publications

Purpose: To develop a new automatic exposure control (AEC) technique based on the contrast-to-noise ratio (CNR) and provide constant lesion detectability.

Methods: Lesion detectability is affected by factors such as image noise, lesion contrast, and lesion size. We performed ROC analysis to assess the relationship between the optimum CNR and the lesion diameter at various levels of lesion contrast.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!