Background: Aim of this study was to compare DNA double-strand breaks (DSBs) in blood lymphocytes of patients undergoing high-pitch helical, low-pitch helical and sequential coronary CT angiography.

Methods And Results: 66 patients were examined with various scan protocols and modes (low-pitch helical scan: 100-120 kV, 320-438 mAs/rot, pitch 0.18-0.39, with or without ECG-pulsing, n=35; prospectively ECG-triggered high-pitch helical scan: 100-120 kV, 320-456 mAs/rotation, pitch 3.2-3.4, n=19; prospectively ECG-triggered sequential scan: 100-120 kV, 150-300 mAs or 320-370 mAs/rotation, n=12) either using a 64-slice or 128-slice dual-source CT or a 128-slice single source CT scanner. Blood samples were obtained before and 30 min after CT and DSBs were analyzed in isolated lymphocytes using γ-H2AX immunofluorescence microscopy. A significant increase of DSBs was measurable 30 min after CTA (range 0.01-0.71/cell). CT induced DSBs showed a significant correlation with the estimated effective dose (ρ=0.90, p<0.00001). Both prospectively ECG-triggered sequential (0.10 DSBs/cell, 176 mGy cm, p<0.00001) and high-pitch helical scan protocols (0.03 DSBs/cell, 109 mGy cm, p<0.00001) led to a significant reduction of median DLP and DSB levels compared to low-pitch helical scans (0.34 DSBs/cell, 828 mGy cm). A reduction of the tube voltage resulted in significantly lower whereas additional calcium scoring resulted in elevated DLP and DNA damages (p<0.05 each).

Conclusion: In coronary CTA, data acquisition protocols have a significant influence on the X-ray induced DSB levels. Using γ-H2AX immunofluorescence microscopy different scan modes in different CT generations can be compared concerning their biological impact.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejrad.2011.11.027DOI Listing

Publication Analysis

Top Keywords

low-pitch helical
12
high-pitch helical
12
scan 100-120
12
dna double-strand
8
double-strand breaks
8
helical scan
8
prospectively ecg-triggered
8
helical
6
x-ray induced
4
induced dna
4

Similar Publications

Natural tristability of a confined helical filament with anisotropic bending rigidities.

Sci Rep

June 2024

Department of Physics, Tamkang University, No. 151 Yingzhuan Rd., Tamsui District, New Taipei City, 251301, Taiwan, ROC.

We find that when and , confining a helical filament with anisotropic bending rigidities within a cylindrical tube of radius R can create a natural tristable status which is consisted of two low-pitch helices and one high-pitch helix, where a helical filament is referred to as a filament that has both an intrinsic curvature ( ) and an intrinsic twist rate ( ). The formation of the tristable status also requires that the filament has a significant difference between two bending rigidities and a large twisting rigidity. The relative heights of two low-pitch helices in a tristable status are close to zero, and the smaller the intrinsic twisting angle, the smaller the difference between these two heights.

View Article and Find Full Text PDF
Article Synopsis
  • The study aimed to analyze how helical pitch and gantry rotation time affect image quality and file size in ultrahigh-resolution photon-counting CT scans of cadaveric specimens.
  • Scans were conducted with varying pitches and rotation times, revealing that while scan duration differed significantly, image quality decreased with high-pitch and short rotation protocols despite consistent dose measures.
  • Findings indicate that careful selection of pitch and rotation time is crucial for optimizing image quality in clinical practices, as longer acquisition times lead to significantly larger raw data sizes, which can limit practical use.
View Article and Find Full Text PDF

Quantitative and qualitative evaluation of three MSCT for high resolution bone imaging.

Eur J Radiol

April 2024

Service D'imagerie Guilloz, CHRU Nancy, Nancy, France; IADI - U1254, Inserm, Université de Lorraine, Nancy, France. Electronic address:

Introduction: Strategies for achieving high resolution varies between manufacturers. In CT, the helical mode with narrow collimation has long been considered as the gold standard for high-resolution imaging. More recently, incremental modes with small dexels and focal spot, have been developed but have not been compared with helical acquisitions under optimal conditions.

View Article and Find Full Text PDF

Low pitch significantly reduces helical artifacts in abdominal CT.

Eur J Radiol

September 2023

Department of Abdominal Imaging, The University of Texas MD Anderson Cancer Center, 1400 Pressler Street, Unit 1473, Houston, TX 77030-4009, United States. Electronic address:

Purpose: High helical pitch scanning minimizes scan times in CT imaging, and thus also minimizes motion artifact and mis-synchronization with contrast bolus. However, high pitch produces helical artifacts that may adversely affect diagnostic image quality. This study aims to determine the severity and incidence of helical artifacts in abdominal CT imaging and their relation to the helical pitch scan parameter.

View Article and Find Full Text PDF

Objective: To measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode.

Methods: A DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!