Evaluation of 25-hydroxy vitamin D assay on the immunodiagnostic systems iSYS analyser.

Ann Clin Biochem

Metabolic Laboratory, Chemical Pathology Directorate, SA Pathology, Box 14, Rundle Post Office, Adelaide SA 5000, Australia.

Published: March 2012

Background: We evaluated the recently released chemiluminescence assay for 25-hydroxy vitamin D (25-OHD) on the Immunodiagnostic Systems iSYS (IDS-iSYS) automated analyser.

Methods: The IDS-iSYS comparison was performed using patient samples previously measured for 25-OHD by a validated liquid chromatography-tandem mass spectrometry (LC-MS/MS) method (n = 119) and an IDS enzyme immunoassy (IDS-EIA) method (n = 64). Limit of detection and limit of quantification were determined from a precision profile. Imprecision was assessed using quality control material and pooled serum. External QAP material (Vitamin D External Quality Assessment Scheme, UK) was analysed to establish inaccuracy. Linearity was assessed by two dilution studies. Cross-reactivity was determined by three serial dilution studies of patient samples with known 25-OHD(2) concentrations.

Results: The IDS-iSYS correlated well with both established methods (iSYS = 1.03LC-MS/MS - 6.53, R(2) = 0.82 and iSYS = 1.07IDS-EIA - 1.61, R(2) = 0.86). Imprecision of the iSYS assay for IDS control material was 13.4% at 32 nmol/L, 10% at 78 nmol/L, 9.4% at 161 nmol/L, and for the pooled material 9.3% at 72 nmol/L and 5.6% at 158 nmol/L. The evaluation found the assay to be highly accurate (IDS-iSYS = 0.93ALTM + 3.79, R(2) = 0.94) and linear (obs(1) = 0.93exp(1) - 5.05, R(2) = 0.99 (P = 0.256); and obs(2) = 0.97exp(2) + 6.07, R(2) = 0.97 (P = 0.654); ALTM, all-laboratory trimmed mean). Cross-reactivity studies demonstrated no significant difference to the calculated total 25-OHD as measured by LC-MS/MS.

Conclusions: Even though the imprecision of the iSYS was found to be greater than that of the LC-MS/MS and EIA methods, the performance characteristics of the IDS-iSYS 25-OHD assay are suitable for routine diagnostic purposes on a high throughput automated analyser.

Download full-text PDF

Source
http://dx.doi.org/10.1258/acb.2011.011018DOI Listing

Publication Analysis

Top Keywords

25-hydroxy vitamin
8
immunodiagnostic systems
8
systems isys
8
patient samples
8
control material
8
dilution studies
8
imprecision isys
8
isys
6
assay
5
ids-isys
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!