Sagittal flexion of the femoral component affects flexion gap and sizing in total knee arthroplasty.

J Arthroplasty

Department of Orthopaedic Surgery, Chiba Rehabilitation Center, 1-45-2 Hondacho, Midori-ku, Chiba, Japan.

Published: June 2012

The purpose of this study was to determine how much sagittal rotation of the femoral component affects the flexion gap and femoral component sizing using a computer-simulation technique. The study comprised 25 knees scheduled for total knee arthroplasty (TKA). The femoral component was positioned at -2°, 0°, 2°, 4°, or 6° of flexion to the anterior femoral cortex, and the resected portion of the posterior medial femoral condyle was measured for 3 total knee systems. The amount of the resected bone of the posterior medial condyle decreased approximately 1 mm for every 2° of additional flexion in all TKA systems. Intentional sagittal flexion of the femoral component by several degrees during TKA can be a useful downsizing technique for the femoral component without excessively increasing the flexion gap.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.arth.2011.10.015DOI Listing

Publication Analysis

Top Keywords

femoral component
24
flexion gap
12
total knee
12
sagittal flexion
8
femoral
8
flexion femoral
8
component flexion
8
knee arthroplasty
8
posterior medial
8
component
6

Similar Publications

The primary weight-bearing structure of the proximal femur, trabecular bone, has a complex three-dimensional architecture that was previously difficult to comprehensively display. This study examined the spatial architecture of trabecular struts in the coronal, sagittal, and horizontal sections of the proximal femur using 21 cases prepared with P45 sectional plasticization. The primary compressive strut (PCS) exhibited a "mushroom-like" shape with upper and lower parts.

View Article and Find Full Text PDF

Background: Many patients with a lower limb socket-suspended prothesis experience socket-related problems, such as pain, chronic skin conditions, and mechanical problems, and as a result, health-related quality of life (HRQoL) is often negatively affected. A bone-anchored prosthesis can overcome these problems and improve HRQoL, but these prostheses have potential downsides as well. A valid and reliable tool to assess potential candidates for surgery concerning a favorable risk-benefit ratio between potential complications related to bone-anchored prostheses and improvements in HRQoL is not available yet.

View Article and Find Full Text PDF

Background: Fixation of distal femoral fractures remains a challenge, and nonunions are common with standard constructs. Far cortical locking (FCL) constructs have been purported to lead to improved fracture-healing as compared with that achieved with traditional locking bridge plates. We sought to test this hypothesis in a comparative effectiveness clinical trial.

View Article and Find Full Text PDF

Diabetes mellitus is a widespread metabolic disorder linked to numerous systemic complications, including adverse effects on skeletal health, such as increased bone fragility and fracture risk. Emerging evidence suggests that high glucose may disrupt the extracellular matrix (ECM) of bone, potentially altering its composition and organization. Collagen, the primary organic component of the ECM, is critical for maintaining structural integrity and biomechanical properties.

View Article and Find Full Text PDF

Instability remains one of the most common indications for revision after total knee arthroplasty. To gain a better understanding of how an implant will perform in vivo and support surgeons in selecting the most appropriate implant design for an individual patient, it is crucial to evaluate the implant constraint within clinically relevant ligament and boundary conditions. Therefore, this study investigated the constraint of three different implant designs (symmetrical implants with and without a post-cam mechanism and an asymmetrical medial-stabilized implant) under anterior-posterior shear forces and internal-external rotation moments at different flexion angles in human cadaveric knees using a six-degrees-of-freedom joint motion simulator.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!