Intrapulmonary effects of setting parameters in portable intrapulmonary percussive ventilation devices.

Respir Care

Acute Neurorespiratory Rehabilitation Unit, Neuromuscular Excellency Centre, Vrije Universiteit Brussel-Inkendaal and Centre for Home Mechanical Ventilation, Ziekenhuis Inkendaal Rehabilitation Hospital, Brussels, Belgium.

Published: May 2012

Background: Despite potential benefits of intrapulmonary percussive ventilation (IPV) in various respiratory diseases, the impact of setting parameters on the mechanical effects produced by IPV in the lungs is unknown. We hypothesized that changing the parameters on IPV would modulate these effects. This in vitro study aimed at comparing the changes in intrapulmonary effects resulting from changes in parameters in 3 portable IPV devices (IMP2, Impulsator, and Pegaso).

Methods: Parameters were set in 72 combinations of frequency (90-250 cycles/min), inspiratory to expiratory (I/E) time ratio (from 1/2 to 3/1), and pressure (10-60 cm H(2)O). Four resulting effects were recorded on a test lung via a pneumotachometer: the expiratory to inspiratory flow ratio (E/I flow ratio), the PEEP, the ventilation, and the percussion. Percussion was assessed by the end-slope of the pressure curve. Analysis of variance was used for data analysis.

Results: E/I flow ratio increased with increasing I/E time ratio (P < .001). The Pegaso produced the lowest E/I flow ratio. PEEP raised 6 cm H(2)O in both IMP2 and Impulsator, and 17 cm H(2)O in the Pegaso with increasing frequency (P < .01), pressure, and I/E time ratio (P < .001). In all devices, ventilation increased with increasing pressure and decreasing frequency (P < .001). Percussion increased with increasing frequency and decreasing I/E time ratio (P < .001), and with increasing pressure when I/E time ratio was 1/1 or less. The Pegaso provided the poorest percussion.

Conclusions: This study suggests that changing the parameters considerably modulates the mechanical effects produced by portable IPV devices in the lungs. Increasing frequency increased PEEP and percussion, but decreased ventilation. Increasing I/E time increased PEEP and E/I flow ratio, and decreased percussion. Finally, increasing pressure increased PEEP and ventilation. The Pegaso produced the highest PEEP, least percussion, and smallest change in E/I flow ratio.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.01441DOI Listing

Publication Analysis

Top Keywords

i/e time
24
flow ratio
24
time ratio
20
e/i flow
20
increased increasing
12
ratio 001
12
increasing frequency
12
increasing pressure
12
increased peep
12
ratio
11

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!