Study Design: Cadaveric Study.

Objective: To compare a fluoroscopic imaging system with computed tomography (CT) and radiographs in detection of spondylolysis and radiation exposure in a cadaver model.

Summary Of Background Data: Lumbar spondylolysis is defined as a defect or fracture of the pars interarticularis and occurs with or without anterior spondylolisthesis. CT scan is the gold standard imaging study for spondylolysis but is limited by the supine position, which may cause reduction of anterolisthesis and by ionizing radiation, which limits the frequency of follow-up scans.

Methods: Thirteen intact cadaveric lumbar spine segments with 26 pars were randomized to be left intact or to undergo simulated fracture using a 1.3 mm oscillating microsurgical saw. Fifteen pars underwent simulated fracture and 11 pars were left intact. Lumbar spine segments were imaged using plain radiographs, multiplanar fluoroscopic imaging, and conventional CT scan. The images were interpreted by 3 observers blinded to the number and location of defects. Radiation exposure and doses were recorded from all imaging units.

Results: Average radiation doses were 0.0025 mSv for each radiograph, 0.23 mSv (low dose) and 0.47 mSv (high dose) for fluoroscopic imaging, and 1.5 mSv for conventional CT imaging (pediatric dose setting). Evaluation of radiographs for spondylolysis had sensitivity of 98% and specificity of 97%. Evaluation using low-dose fluoroscopic images, high-dose fluoroscopic images, and CT scan images correctly identified the status of all pars based on multiplanar images; sensitivity and specificity were 100%. Kappa analysis demonstrated a value of 0.89 for radiographic interpretation indicating excellent agreement. Kappa values describing agreement for image interpretation for fluoroscopic imaging and CT scan were equal to 1.0, representing perfect agreement.

Conclusions: Three-dimensional fluoroscopic imaging provides comparable diagnostic imaging with CT scan in an experimental cadaveric model of spondylolysis using up to 85% less radiation than conventional CT scan.

Download full-text PDF

Source
http://dx.doi.org/10.1097/BSD.0b013e318228bcccDOI Listing

Publication Analysis

Top Keywords

fluoroscopic imaging
20
lumbar spine
12
imaging
10
imaging system
8
cadaveric lumbar
8
radiation exposure
8
fracture pars
8
spine segments
8
left intact
8
simulated fracture
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!