The forced oscillation technique (FOT) is a non-invasive technique to monitor airway obstruction in those with asthma. The aim of this study was to design and validate a system to use FOT during sleep, both with and without bi-level positive airway pressure (BPAP), and to separate upper airway resistance from lower. 8 Hz pressure oscillations were supplied, over which the subject breathed, pressure and flow measurements were then used to calculate impedance. A phase-shift induced by the pressure transducer tubing was characterized, and FOT resistance was compared to steady flow resistance both with and without BPAP. A Millar catheter was used to measure pressure at the epiglottis, allowing the separation of upper from lower airway resistance. A phase shift of -0.010 s was calculated for the pressure transducer tubing, and the average error between FOT and steady flow resistance was -0.2 ± 0.2 cmH₂O/L/s without BPAP and 0.4 ± 0.2 cmH₂O/L/s with BPAP. The system was tested on three subjects, one healthy, one with obstructive sleep apnea, and one with asthma. The FOT was well tolerated and resistance was separated into upper and lower airway components. This setup is suitable for monitoring both upper and lower airway obstruction during sleep in those with and without asthma.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3387574PMC
http://dx.doi.org/10.1007/s10439-011-0470-7DOI Listing

Publication Analysis

Top Keywords

upper lower
16
lower airway
16
airway resistance
12
forced oscillation
8
oscillation technique
8
airway obstruction
8
pressure transducer
8
transducer tubing
8
steady flow
8
flow resistance
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!