Philips DigitalDiagnost, a digital radiographic system mounting flat panel detector (FPD), can display dose area data (DiDi dose) calculated by examination parameters. We evaluated its fundamental characteristics and compared the values of DiDi dose andactual measured data obtained by the area dose product meter (PD-4100L). Tendency of varied values of mAs, X-ray tube values and exposure area from both the area dose product meter and the DiDi dose were coincided. Further, in clinical images of chest PA 100 cases, chest lateral 50 cases and abdomen stand 25 cases, the determination coefficient was overly high as R(2)=0.99. Based on these results, it is clear that the DiDi dose can be treated the same as the area dose product meter. Under increasing of patient X-ray exposure dose is a concern in digital general radiography, this research indicates that maximum values of histogram obtained by DiDi dose contributes dose awareness for radiographer.