Purpose: The authors evaluated the diagnostic accuracy of second-generation dual-source (DSCT) computed tomography coronary angiography (CTCA) with iterative reconstructions for detecting obstructive coronary artery disease (CAD).
Materials And Methods: Between June 2010 and February 2011, we enrolled 160 patients (85 men; mean age 61.2±11.6 years) with suspected CAD. All patients underwent CTCA and conventional coronary angiography (CCA). For the CTCA scan (Definition Flash, Siemens), we use prospective tube current modulation and 70-100 ml of iodinated contrast material (Iomeprol 400 mgI/ ml, Bracco). Data sets were reconstructed with iterative reconstruction algorithm (IRIS, Siemens). CTCA and CCA reports were used to evaluate accuracy using the threshold for significant stenosis at ≥50% and ≥70%, respectively.
Results: No patient was excluded from the analysis. Heart rate was 64.3±11.9 bpm and radiation dose was 7.2±2.1 mSv. Disease prevalence was 30% (48/160). Sensitivity, specificity and positive and negative predictive values of CTCA in detecting significant stenosis were 90.1%, 93.3%, 53.2% and 99.1% (per segment), 97.5%, 91.2%, 61.4% and 99.6% (per vessel) and 100%, 83%, 71.6% and 100% (per patient), respectively. Positive and negative likelihood ratios at the per-patient level were 5.89 and 0.0, respectively.
Conclusions: CTCA with second-generation DSCT in the real clinical world shows a diagnostic performance comparable with previously reported validation studies. The excellent negative predictive value and likelihood ratio make CTCA a first-line noninvasive method for diagnosing obstructive CAD.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s11547-011-0754-x | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!