  • This study aimed to compare the effectiveness of two hydration measurement instruments, the Corneometer 825® and the Skicon-200 EX®, particularly focusing on their calibration and accuracy in measuring stratum corneum hydration.
  • In both controlled lab settings and real skin tests, researchers found that while capacitance measurements can assess deeper skin layers and aren't affected by electrolytes, conductance measurements are limited to shallower depths and influenced by saline content.
  • The results indicated a strong correlation between the two methods across different skin hydratation levels, suggesting both instruments are reliable for estimating skin hydration despite some limitations in sensitivity.

Article Abstract

Background: Measurement of stratum corneum (SC) hydration often involves the use of commercial instruments. The aim of this study was to compare and validate two recent instruments: the Corneometer 825® (digital probe) and the Skicon-200 EX®.

Methods: In vitro calibration was carried out on filter pads using different solvents, measurements over different layers of plastic foils, and evaluation of desorption kinetics. In vivo measurements were carried out on skin sites covering a range of very dry to well-hydrated skin areas.

Results: Conductance measurements are influenced by electrolytes while capacitance measurements are not. Dielectric constant of the solvents influences the values of both instruments (r respectively 0.92 and 0.99). The capacitance method carries information from deeper layers (up to 45 μm) compared with the conductance instrument (up to 15 μm). Desorption experiments show a strong relation between the amount of water and respectively the capacitance and the conductance values. The in vivo experiments revealed a strong relation between the two methods (r = 0.97). Sensitivity of the capacitance method is limited for the highest hydration values.

Conclusions: Both instruments allow a certain calibration and both give good estimates of the SC hydration.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1600-0846.2011.00573.xDOI Listing

Publication Analysis

Top Keywords

capacitance method
12
stratum corneum
8
corneometer 825®
8
strong relation
8
capacitance
5
hydration
4
hydration measurements
4
measurements stratum
4
corneum comparison
4
comparison capacitance
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!