The purpose of this study was to investigate the effect of parameter changes that can potentially lead to unreliable measurements in fat quantification. Chemical shift imaging was performed using spoiled gradient echo sequences with systematic variations in the following: two-dimensional/three-dimensional sequence, number of echoes, delta echo time, fractional echo factor, slice thickness, repetition time, flip angle, bandwidth, matrix size, flow compensation and field strength. Results indicated no significant (or significant but small) changes in fat fraction with parameter. The significant changes can be attributed to the known effects of T1 bias and two forms of noise bias.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3254829PMC
http://dx.doi.org/10.1016/j.mri.2011.09.011DOI Listing

Publication Analysis

Top Keywords

fat quantification
8
quantification chemical
8
chemical shift
8
shift imaging
8
parameter changes
8
robustness fat
4
imaging purpose
4
purpose study
4
study investigate
4
investigate parameter
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!