A cryogenic loading system for introducing H(2) and D(2) into the diamond anvil cell has been designed and constructed. The integration of pressure loading mechanism, ruby fluorescence spectrometer, and microscope camera allows for in situ tuning and calibrating the pressure. The performance of the system has been demonstrated by successful synthesis of hydride and deuteride of transition metal and rare earth metal. Our cryogenic methodology features facile start-over of loading and in situ electrical resistance measurement of as-synthesized sample.