The potential for peri-implant bone yielding and subsequent loosening of Ilizarov ring-wire external fixation systems was investigated using non-linear finite element (FE) analyses. A strain-based plasticity model was employed to simulate bone yielding. FE models also incorporated contact behavior at the wire-bone interface, orthotropic elasticity, and periosteal-endosteal variation of bone properties. These simulations were used to determine the extent and location of yielding with change in age-related bone structure and properties for the bone-Ilizarov construct at the tibial midshaft. At critical wire-bone interfaces, the predicted volume of yielded bone with four wires (on either side of the fracture) was ∼40% of that with two wires. Old-aged cases showed considerably greater bone yielding at the wire-bone interface than young cases (1.7-2.2 times greater volumes of yielded bone). The volume of yielded bone at all wire-bone interfaces decreased with an increase in wire pre-tension. The absence of continuous through-thickness yielding offers an explanation for the clinical observation that Ilizarov ring-wire fixation can provide stable fracture fixation even in bone with high porosity.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jor.21587DOI Listing

Publication Analysis

Top Keywords

ilizarov ring-wire
12
bone yielding
12
yielded bone
12
bone
9
loosening ilizarov
8
ring-wire external
8
wire-bone interface
8
wire-bone interfaces
8
volume yielded
8
yielding
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!