Temperature rise during photo-polymerization for onlay luting.

Am J Dent

School of Dentistry, University of Geneva, Rue Barthélemy-Menn 19, CH-1205 Geneva, Switzerland.

Published: August 2011

Purposes: (1) To measure the temperature rise during long-time irradiation needed to lute adhesive indirect restorations, with one halogen and five LED high-power lamps, in a simulated oral environment, and (2) to propose a technique to reduce heat.

Methods: Temperature within the pulp chamber of an extracted human molar in a 35 degrees C water bath, before and after restoration with a ceramic onlay, was continuously recorded during 3 x 120 seconds of irradiation for different curing regimens, with and without cooling by compressed air, water and water spray. Temperature rise in a thermo-insulated well was also measured during 120 seconds of irradiation with LEDemetron II and TuTu.

Results: Maximal temperature rise of 6.2 degrees C/7.7 degrees C on intact/onlay restored tooth was found after 3 x 120 seconds of irradiation. Compressed air application reduced the temperature by 4 degrees C for LEDemetron II, while water spray increased the temperature by 15.1 degrees C for TuTu. Temperature decreased with irradiation distance, except for LEDemetron II. Air was heated up to 65 degrees C after 120 seconds of irradiation in an insulated well with TuTu.

Download full-text PDF

Source

Publication Analysis

Top Keywords

temperature rise
16
120 seconds
16
seconds irradiation
16
temperature
8
compressed air
8
water spray
8
irradiation
6
degrees
6
rise photo-polymerization
4
photo-polymerization onlay
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!