Purpose: The aim of this study is to compare bipolar radiofrequency-driven vessel sealer, bipolar electrocautery, polyglactin 910 sutures, and endoclips in appendiceal stump closure with respect to operative time, appendiceal stump strength, and inflammation in a rat appendiceal model.

Methods: Forty-eight Wistar-Albino rats, which had previously created appendicitis, were divided into 2 (groups A and B). Each group was further subdivided into 4 subgroups (AL, ABPC, AC, AS, BL, BBPC, BC, and BS) each containing 6 rats. An appendectomy was performed, and the stump was closed by bipolar radiofrequency-driven vessel sealer in the L, bipolar electrocautery in the BPC, endoclips in the C, and ligatures in the S subgroups. Cecum bursting pressures were determined instantly after the operation in group A and on the seventh postoperative day in group B. All operative times were measured. Appendices and appendiceal stumps underwent histopathologic examination. Statistical analyses were performed with Kruskal-Wallis and Mann-Whitney U tests. P < .05 was considered significant.

Results: Bursting pressures of the subgroups were comparable on postoperative immediate period and day 7. Operative times were significantly shorter in the L and BPC subgroups in A and B. Histopathologic examination showed that the inflammation scores were similar in group A. In group B, inflammation parameters were also similar except the necrosis status, which was found to be decreased in BL compared with BC and BS. Necrosis status was significantly lower, and lymphocyte quantity was significantly higher in BL and BBPC compared with AL and ABPC.

Conclusions: Bipolar radiofrequency-driven vessel sealer and bipolar electrocautery achieve safe stump closure with satisfactory bursting pressure values in an experimental rat appendicitis model. Decreased operative time and unimpaired healing are encouraging.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jpedsurg.2011.04.062DOI Listing

Publication Analysis

Top Keywords

bipolar radiofrequency-driven
12
radiofrequency-driven vessel
12
vessel sealer
12
sealer bipolar
12
bipolar electrocautery
12
rat appendicitis
8
appendicitis model
8
appendiceal stump
8
stump closure
8
operative time
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!