Purpose: To evaluate the accuracy of 3 spectral-domain (SD) optical coherence tomography (OCT) devices (Topcon 3D-1000 [Topcon]; Cirrus HD [Carl Zeiss Meditec, Inc], and Spectralis OCT [Heidelberg Engineering]) before and after mydriasis for the diagnosis of diabetic macular edema.

Design: Cross-sectional study.

Methods: Sixty-two eyes of 62 consecutive patients with diabetes without recent loss of vision referred for retinal control were assessed. Two scans were performed for each SD OCT instrument. Central retinal thickness was measured before and after pupil dilation. Pupil dynamic was studied using pMetrics pupillometer (iVIS Technologies), and lens opacity was measured by Pentacam densitometry (Oculus). The diagnostic accuracy of SD OCT devices was assessed by sensitivity, specificity, and area under the receiver operating characteristic curve. Logistic regression analysis was used to assess the effect of pupil size and lens opacity on the reliability of SD OCT in the acquisition of adequate images.

Results: The area under the receiver operating characteristic curve for the Topcon 3D OCT device was 0.84, that for the Cirrus HD OCT device was 0.93, and that for the Spectralis OCT device was 0.91. Significant differences in area under the receiver operating characteristic curve before and after pupillary dilatation were not found. Sensitivity and specificity associated with the cutoff value for the best performance were 82% and 74% for the Topcon 3D OCT device, 90% and 87% for the Cirrus HD OCT device, and 90% and 84% for the Spectralis OCT device, respectively. The Topcon 3D OCT device had an 11.3% segmentation algorithm failure rate for the central millimeter of the fovea, and the nuclear lens density was significantly greater in these eyes than in those without failure (17.1 ± 1.1 mm vs 10.4 ± 0.2 mm; P < .05).

Conclusions: SD OCT is a useful tool to detect and to measure diabetic macular edema without the need for pupil dilatation.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajo.2011.08.008DOI Listing

Publication Analysis

Top Keywords

oct device
28
oct
13
diabetic macular
12
spectralis oct
12
area receiver
12
receiver operating
12
operating characteristic
12
characteristic curve
12
topcon oct
12
spectral-domain optical
8

Similar Publications

Esophago-jejunal anastomoses fistula could be mortal. Currently there is a wide therapeutic measure ranging from conservative management, endoscopic therapy and surgery. Endoscopic management has been positioned above other strategies due to minimal invasion which improves survival and reduces mortality.

View Article and Find Full Text PDF

Objective: To investigate the effects of bulk-fill, resin-based composite types (high or low viscosity) on the internal adaptation of Class V restorations.

Study Design: Experimental study. Place and Duration of the Study: Hefei Stomatological Hospital, Hefei, China, from October 2022 to December 2023.

View Article and Find Full Text PDF

Background/ Aims: To analyze the longitudinal change in Bruch's membrane opening minimal rim width (BMO-MRW) and peripapillary retinal nerve fiber layer (pRNFL) thickness using optical coherence tomography (OCT) after implantation of a PRESERFLO® microshunt for surgical glaucoma management in adult glaucoma patients.

Methods: Retrospective data analysis of 59 eyes of 59 participants undergoing implantation of a PRESERFLO microshunt between 2019 and 2022 at a tertiary center for glaucoma management. Surgical management included primary temporary occlusion of the glaucoma shunt to prevent early hypotony.

View Article and Find Full Text PDF

Extensive research on retinal layer segmentation (RLS) using deep learning (DL) is mostly approaching a performance plateau, primarily due to reliance on structural information alone. To address the present situation, we conduct the first study on the impact of multi-spectral information (MSI) on RLS. Our experimental results show that incorporating MSI significantly improves segmentation accuracy for retinal layer optical coherence tomography (OCT) images.

View Article and Find Full Text PDF

The impact of various neurodegenerative diseases on the retina has been investigated in recent years using optical coherence tomography (OCT). Epilepsy, classified as a neurodegenerative disorder, has been indicated to affect the structural integrity of the retina. Moreover, there is ongoing debate regarding the relative contribution of disease pathogenesis and the consumption of anti-epileptic drugs (AEDs) to these retinal changes.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!