We present the x-ray optical design of the soft x-ray materials science instrument at the Linac Coherent Light Source, consisting of a varied line-spaced grating monochromator and Kirkpatrick-Baez refocusing optics. Results from the commissioning of the monochromator are shown. A resolving power of 3000 was achieved, which is within a factor of two of the design goal.