Background: Magnetic resonance imaging (MRI) is avoided in most patients with implanted cardiac devices because of safety concerns.

Objective: To define the safety of a protocol for MRI at the commonly used magnetic strength of 1.5 T in patients with implanted cardiac devices.

Design: Prospective nonrandomized trial. (ClinicalTrials.gov registration number: NCT01130896) SETTING: One center in the United States (94% of examinations) and one in Israel.

Patients: 438 patients with devices (54% with pacemakers and 46% with defibrillators) who underwent 555 MRI studies.

Intervention: Pacing mode was changed to asynchronous for pacemaker-dependent patients and to demand for others. Tachyarrhythmia functions were disabled. Blood pressure, electrocardiography, oximetry, and symptoms were monitored by a nurse with experience in cardiac life support and device programming who had immediate backup from an electrophysiologist.

Measurements: Activation or inhibition of pacing, symptoms, and device variables.

Results: In 3 patients (0.7% [95% CI, 0% to 1.5%]), the device reverted to a transient back-up programming mode without long-term effects. Right ventricular (RV) sensing (median change, 0 mV [interquartile range {IQR}, -0.7 to 0 V]) and atrial and right and left ventricular lead impedances (median change, -2 Ω [IQR, -13 to 0 Ω], -4 Ω [IQR, -16 to 0 Ω], and -11 Ω [IQR, -40 to 0 Ω], respectively) were reduced immediately after MRI. At long-term follow-up (61% of patients), decreased RV sensing (median, 0 mV, [IQR, -1.1 to 0.3 mV]), decreased RV lead impedance (median, -3 Ω, [IQR, -29 to 15 Ω]), increased RV capture threshold (median, 0 V, IQR, [0 to 0.2 Ω]), and decreased battery voltage (median, -0.01 V, IQR, -0.04 to 0 V) were noted. The observed changes did not require device revision or reprogramming.

Limitations: Not all available cardiac devices have been tested. Long-term in-person or telephone follow-up was unavailable in 43 patients (10%), and some data were missing. Those with missing long-term capture threshold data had higher baseline right atrial and right ventricular capture thresholds and were more likely to have undergone thoracic imaging. Defibrillation threshold testing and random assignment to a control group were not performed.

Conclusion: With appropriate precautions, MRI can be done safely in patients with selected cardiac devices. Because changes in device variables and programming may occur, electrophysiologic monitoring during MRI is essential.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4337840PMC
http://dx.doi.org/10.7326/0003-4819-155-7-201110040-00004DOI Listing

Publication Analysis

Top Keywords

cardiac devices
16
patients implanted
12
implanted cardiac
12
patients
9
magnetic resonance
8
resonance imaging
8
sensing median
8
median change
8
median [iqr
8
capture threshold
8

Similar Publications

Purpose: Atrial fibrillation (AF) is the most common chronic cardiac arrhythmia that increases the risk of stroke, primarily due to thrombus formation in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) devices offer an alternative to oral anticoagulation for stroke prevention. However, the complex and variable anatomy of the LAA presents significant challenges to device design and deployment.

View Article and Find Full Text PDF

Description of current status of implementation and management of cardiac arrest in China.

Sci Rep

January 2025

Emergency Department, The State Key Laboratory for Complex, Severe and Rare Diseases, Peking Union Medical College Hospital, Chinese Academy of Medical Science and Peking Union Medical College, Beijing, China.

Variation in the incidence, survival rate and factors associated with survival after cardiac arrest in China is reported. Some studies have tried to fill the knowledge gap regarding the epidemiology of cardiac arrest in China but were unable to identify reasons for the reported differences. Therefore, the purpose of this study was to describe Chinese management of cardiac arrest, particularly from the perspective of compression, ventilation, monitoring, treatment, and extracorporeal cardiopulmonary resuscitation.

View Article and Find Full Text PDF

Introduction: Altered vascular microcirculation is recognized as a risk factor for anastomotic leakage (AL) in colorectal surgery. However, few studies evaluated its impact on AL using different devices, with heterogeneous results. The present study reported the initial experience measuring gut microcirculatory density and flow with the aid of incidence dark-field (IDF) videomicroscopy (Cytocam, Braedius, Amsterdam, The Netherlands) comparing its operative outcome using a propensity score matching (PSM) model based on age, gender, and Charlson Comorbidity Index (CCI).

View Article and Find Full Text PDF

A young man in his 30s presented to us with multiple episodes of syncope and exertional dyspnoea for the last 2 weeks. He was diagnosed with squamous cell carcinoma of the lower one-third of the oesophagus in 2021 for which he was treated with neoadjuvant chemoradiotherapy, followed by McKeown oesophagectomy. At 2-year follow-up, he had developed a soft tissue swelling in the scalp, which was diagnosed as a tumour recurrence and radiotherapy was initiated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!