Purpose: The biological responses to mesh in vivo have been evaluated in some papers, but the in vivo condition of mesh and plugs have not been sufficiently evaluated. This study evaluated the endoscopic observations and histological assessments of mesh plugs using swine models.
Methods: An artificial abdominal hernia was established in the porcine abdomen, and repaired using three different sizes of two types of plug, Proloop (ATRIUM Medical Corporation, Hudson, NH, USA) or Perfix (BARD Medical Division, Covington, GA, USA). The in vivo conditions of each plug were periodically observed using a laparoscope. Moreover, a histological evaluation of the plugs was performed 3 months after implantation.
Results: The laparoscopic observation revealed that inversion of the plugs occurred in 10 out of 18 cases repaired with Perfix, while no case repaired with Proloop inverted. The large and medium sizes of Perfix plugs were inclined by an average of more than 30°. In addition, the triangular shape of Perfix plugs was broken and the vertical/horizontal ratio was enlarged during the observation period, while Proloop plugs shrank both vertically and horizontally. The inflammatory cell count was significantly lower within the Proloop plugs than within Perfix plugs.
Conclusion: Proloop plugs are apparently superior because they are stable even 3 months after implantation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00595-010-4520-9 | DOI Listing |
Surg Today
November 2011
Department of Cancer and Thoracic Surgery, Graduate School of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University, 2-5-1 Shikata-cho, Kita-ku, Okayama, 700-8558, Japan.
Purpose: The biological responses to mesh in vivo have been evaluated in some papers, but the in vivo condition of mesh and plugs have not been sufficiently evaluated. This study evaluated the endoscopic observations and histological assessments of mesh plugs using swine models.
Methods: An artificial abdominal hernia was established in the porcine abdomen, and repaired using three different sizes of two types of plug, Proloop (ATRIUM Medical Corporation, Hudson, NH, USA) or Perfix (BARD Medical Division, Covington, GA, USA).
Hernia
October 2009
Department of Surgery, Derriford Hospital, Plymouth PL6 8DH, UK.
Introduction: Mesh plug is an established and effective method for repair of inguinal hernia. The ProLoop plug (Atrium) is a recently developed mesh plug with a novel configuration, which may offer advantages over the standard Prefix plug (Bard) or the Lichtenstein repair. This two-centre double-blinded randomised control trial assessed the short- and medium-term outcomes, comparing the above three methods.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!