Background: Surveys indicate that up to 90% of general surgeons and gynecologists use monopolar radiofrequency during laparoscopy and 18% have experienced visceral burns. Monopolar electrosurgery compared with other energy sources is associated with unique characteristics and inherent risks and complications caused by inadvertent direct or capacitive coupling or insulation failure of instruments. These dangers become particularly important with the reemergence of single-port laparoscopy, which requires close proximity and crossing of multiple intraabdominal instruments outside the surgeon's field of view.

Study Objectives: To determine the effects of monopolar electrosurgery on various tissues/organs during simulated single-port laparoscopic surgery in vitro and in vivo.

Design: Simulation in a dry laboratory with fresh sheep liver, pig bowel and bowel in an anesthetized dog (Canadian Classification II-3).

Setting: University-affiliated teaching hospital and animal facilities.

Measurements And Main Results: We used Valleylab Force 2 and FX electrosurgical generators at clinically used power outputs of 40 to 60 watts, and both high- and low-voltage (coagulation and cut) waveforms and commercially-available single-port devices. The effect on tissue was recorded by pictures and video camera and graded visually and histologically with hematoxylin and eosin stains. During activation of any standard monopolar laparoscopic instrument (scissors, coagulating electrode, etc), capacitive coupled currents resulting in visible tissue burn (blanching) caused by other adjacent cold instrument (graspers, etc) including metallic suction-irrigation cannulas and the laparoscope itself were noted. Histopathologic study confirmed transmural thermal damage extending to the mucosa of small bowel, even in the presence of mild serosa blanching. With prolonged activation of the electrosurgical generator, the capacitive coupled corona discharge burned the insulation and caused rapid insulation breakdown of the electrode instrument resulting in direct coupling (sparking, arcing) to adjacent cold instruments and more severe burning to the contacted tissue/organ.

Conclusions: During single-port laparoscopy and use of monopolar radiofrequency, the proximity and crossing of multiple instruments generate capacitive or direct coupled currents, which may cause visceral burns.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jmig.2011.07.015DOI Listing

Publication Analysis

Top Keywords

monopolar electrosurgery
12
single-port laparoscopy
12
monopolar radiofrequency
8
visceral burns
8
proximity crossing
8
crossing multiple
8
capacitive coupled
8
coupled currents
8
adjacent cold
8
monopolar
6

Similar Publications

Rare case of cochlear implant damaged by nearby use of monopolar electrosurgery.

Am J Otolaryngol

December 2024

Department of Otolaryngology-Head & Neck Surgery, University of California San Francisco, San Francisco, CA, United States of America. Electronic address:

Objectives: To present a rare case of a cochlear implant (CI) damaged by nearby use of monopolar electrosurgery.

Patient: A 38-year-old man with a right-sided CI reported that his implant had stopped producing sound immediately after his meningioma resection.

Interventions: Right pterional craniotomy with use of monopolar electrosurgery.

View Article and Find Full Text PDF

Surgical cutting with electrosurgical tools facilitates tissue dissection and vessel sealing, preventing blood loss. The extent of tissue necrosis due to temperature elevations is dependent on the cutting technique, device design, coating properties and power settings, but the influence of these parameters is not fully understood. Here we conduct a comprehensive comparative analysis of thermal damage comparing (1) manual user-controlled and robotic electrosurgical cutting approaches for (2) varying electrodes and coatings, and power settings.

View Article and Find Full Text PDF

Hysteroscopic surgery requires a balance of continuous controlled irrigation and aspiration to distend the endometrial cavity to a degree that provides the clear and stable visual environment necessary for diagnostic and therapeutic procedures. Whereas the preferred distending solution should be isotonic and isonatremic, radiofrequency (RF) electrosurgery with monopolar instrumentation can only be performed with non-ionic (hyponatremic) solutions. Absorption of as little as 500 mL and certainly more than 1000 mL of non-ionic solutions can result in fluid overload and/or dilutional hyponatremia with potentially serious adverse effects under certain conditions and patient characteristics.

View Article and Find Full Text PDF

Comparison of two different electrosurgical devices in reduction mammoplasty: Monopolar electrocautery versus plasmakinetic cautery.

World J Surg

August 2024

Department of Plastic Reconstructive and Aesthetic Surgery, University of Health Sciences, Konya City Hospital, Konya, Turkey.

Background: Many devices are used for dissection and hemostasis during reduction mammoplasty. While one of the most common methods is monopolar electrocautery, tissue damage due to thermal spread remains a controversial topic. New devices have been designed to minimize this effect.

View Article and Find Full Text PDF

Introduction: Operating room fires can have devastating consequences and as such must be prevented. There exists a paucity of literature requiring further elucidation regarding manufacturer recommendations of a predefined waiting period prior to patient draping after using alcohol-based surgical antiseptics, in order to reduce the risk of operating room fires.

Methods: This was further investigated by exposing two common alcohol-based surgical antiseptics to electrosurgery and open flames at various power settings and time intervals in an ex vivo porcine model.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!