Background: Respiratory tract infections are the most common cause of hospitalization in infants and young children and are typically caused by viral or, less commonly, bacterial pathogens. Existing non-molecular diagnostic methods have several drawbacks such as limited sensitivity, long turn-a-round time and limited number of pathogens that can be detected.

Objectives: Nucleic acid amplification methods can increase sensitivity and enable the initiation of appropriate interventions without delay. Broad-spectrum detection and identification circumvent the use of individual diagnostic DNA or RNA based assays. At present, several commercial assays are available for broad-spectrum detection.

Study Design: We compared the performance of the xTAG Respiratory Viral Panel (RVP) (Luminex Molecular Diagnostics, Toronto, Canada) with that of the Respifinder (Pathofinder, Maastricht, Netherlands) for 9 external quality assurance (EQA) panels (QCMD, Scotland) consisting of a total of 106 EQA samples.

Results: Both the RVP and the Respifinder assay have an excellent specificity. Sensitivity was 33% and 78% for the RVP and the Respifinder assay, respectively. For both assays, sensitivity was low for weak positive samples.

Discussion: The results of our study seem to indicate a better sensitivity for the Respifinder. Analysis of patient samples is necessary to evaluate the clinical performance.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7185546PMC
http://dx.doi.org/10.1016/j.jcv.2011.08.017DOI Listing

Publication Analysis

Top Keywords

respifinder assay
12
respiratory tract
8
tract infections
8
xtag respiratory
8
respiratory viral
8
viral panel
8
rvp respifinder
8
respifinder
5
sensitivity
5
timely diagnosis
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!