Purpose: To compare the increase in intraocular pressure (IOP) during corneal flap preparation in porcine eyes when using a femtosecond laser or a mechanical microkeratome.
Setting: The present study was conducted at a university hospital and a private clinic.
Methods: The vitreous IOP was directly measured with a cannula through the optic nerve in 24 porcine globes (n = 12 for each device). In the first procedure (worst-case procedure), the eye interface was lowered against the globe until abortion of the docking maneuver when using the IntraLase femtosecond laser (Abbott Medical Optics, Santa Ana, CA) or the suction ring was pressed very firmly against the globe when using the Amadeus microkeratome (Ziemer Ophthalmic Systems AG, Port, Switzerland). In the second procedure (regular procedure), a normal lamellar flap creation was performed using each of these devices.
Results: During the worst-case procedure, a maximum IOP of 260 ± 53 mm Hg (mean ± SD) was reached with the IntraLase and 318 ± 59 mm Hg was reached with the Amadeus. During the regular procedure, the maximum IOP increased to 135 ± 16 mm Hg when using the IntraLase femtosecond laser and to 152 ± 24 mm Hg with the Amadeus microkeratome.
Conclusions: In the worst-case procedure, the maximum IOP levels were lower in the IntraLase group. No significant difference in maximum IOP levels could be obtained during the regular procedure. The duration of the suction phase is longer when using the IntraLase femtosecond laser.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/ICO.0b013e318212110a | DOI Listing |
Heliyon
January 2025
Laser and Plasma Research Institute, Shahid Beheshti University, Tehran, Iran.
The MKN45 cell line, a type of gastric cancer cell, exhibits resistance to chemotherapy agents through various mechanisms. Curcumin and noscapine, two plant-derived anticancer compounds, exhibit selective cytotoxicity towards cancer cells. However, their bioavailability is poor both in vitro and in vivo.
View Article and Find Full Text PDFLasers Surg Med
January 2025
ViaLase Inc., Aliso Viejo, California, USA.
Objectives: Femtosecond laser image guided high precision trabeculotomy (FLigHT) is a novel open-angle glaucoma treatment. The procedure non-invasively creates aqueous humor (AH) drainage channel from the anterior chamber (AC) to Schlemm's canal (SC) through the trabecular meshwork (TM) to decrease intraocular pressure (IOP). The purpose of this study was to develop a 3D finite element model (FEM) of the FLigHT procedure and to simulate clinical results for different drainage channel cross-sectional areas.
View Article and Find Full Text PDFJ Prosthodont Res
January 2025
Department of Gerodontology and Oral Rehabilitation, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.
Purpose: This study investigated the effects of femtosecond laser (FL) irradiation on the surface roughness and shear bond strength of high-translucency zirconia (6 mol% yttria-partially stabilized zirconia [6Y-PSZ]) and lithium disilicate (LiSiO) glass ceramics.
Methods: Fully sintered square-shaped specimens of 6Y-PSZ (7 groups; 20 specimens/group) and LiSiO (8 groups; 20 specimens/group) were surface-treated via sandblasting (50-μm alumina sand or glass beads) or FL irradiation (20- or 40-μm dot or cross-line patterns) or using Monobond Etch & Prime (Ivoclar Vivadent AG; only for LiSiO specimens). The surface roughness (arithmetic average [Sa] and developed interfacial area ratio [Sdr]) and shear bond strength after 24 h and 10,000 thermal cycles were measured and statistically analyzed.
Sci Rep
January 2025
Department of Physics, University of Ottawa, Ottawa, ON, K1N 6N5, Canada.
We report a nonlinear terahertz (THz) detection device based on a metallic bull's-eye plasmonic antenna. The antenna, fabricated with femtosecond laser direct writing and deposited on a nonlinear gallium phosphide (GaP) crystal, focuses incoming THz waveforms within the sub-wavelength bull's eye region to locally enhance the THz field. Additionally, the plasmonic structure minimizes diffraction effects allowing a relatively long interaction length between the transmitted THz field and the co-propagating near-infrared gating pulse used in an electro-optic sampling configuration.
View Article and Find Full Text PDFSci Rep
January 2025
Laser Research Center, Vilnius University, Saulėtekio Avenue 10, LT-10223, Vilnius, Lithuania.
We present a comparative experimental study of supercontinuum generation in undoped scintillator crystals: bismuth germanate (BGO), yttrium orthosilicate (YSO), lutetium oxyorthosilicate (LSO), lutetium yttrium oxyorthosilicate (LYSO) and gadolinium gallium garnet (GGG), pumped by 180 fs fundamental harmonic pulses of an amplified Yb:KGW laser. In addition to these materials, experiments in yttrium aluminium garnet (YAG), potassium gadolinium tungstate (KGW) and lithium tantalate (LT) were performed under identical experimental settings (focusing geometry and sample thickness), which served for straightforward comparison of supercontinuum generation performances. The threshold and optimal (that produces optimized red-shifted spectral extent) pump pulse energies for supercontinuum generation were evaluated from detailed measurements of spectral broadening dynamics.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!