Objective: To evaluate three commercial typhoid rapid antibody tests for Salmonella Typhi antibodies in patients suspected of having typhoid fever in Mpumalanga, South Africa, and Moshi, United Republic of Tanzania.
Methods: The diagnostic accuracy of Cromotest(®) (semiquantitative slide agglutination and single tube Widal test), TUBEX(®) and Typhidot(®) was assessed against that of blood culture. Performance was modelled for scenarios with pretest probabilities of 5% and 50%.
Findings: In total 92 patients enrolled: 53 (57.6%) from South Africa and 39 (42.4%) from the United Republic of Tanzania. Salmonella Typhi was isolated from the blood of 28 (30.4%) patients. The semiquantitative slide agglutination and single-tube Widal tests had positive predictive values (PPVs) of 25.0% (95% confidence interval, CI: 0.6-80.6) and 20.0% (95% CI: 2.5-55.6), respectively. The newer typhoid rapid antibody tests had comparable PPVs: TUBEX(®), 54.1% (95% CI: 36.9-70.5); Typhidot(®) IgM, 56.7% (95% CI: 37.4-74.5); and Typhidot(®) IgG, 54.3% (95% CI: 36.6-71.2). For a pretest probability of 5%, PPVs were: TUBEX(®), 11.0% (95% CI: 6.6-17.9); Typhidot(®) IgM, 9.1% (95% CI: 5.8-14.0); and Typhidot(®) IgG, 11.0% (6.3-18.4). For a pretest probability of 50%, PPVs were: TUBEX(®), 70.2% (95% CI: 57.3-80.5); Typhidot(®) IgM, 65.6% (95% CI: 54.0-75.6); and Typhidot(®) IgG, 70.0% (95% CI: 56.0-81.1).
Conclusion: Semiquantitative slide agglutination and single-tube Widal tests performed poorly. TUBEX(®) and Typhidot(®) may be suitable when pretest probability is high and blood cultures are unavailable, but their performance does not justify deployment in routine care settings in sub-Saharan Africa.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3165980 | PMC |
http://dx.doi.org/10.2471/BLT.11.087627 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!