Background: Asymmetric dimethylarginine (ADMA), an endogenous competitive inhibitor of nitric oxide synthase plays an important role in endothelial dysfunction processes. Recent studies have linked high ADMA levels with several pathological conditions. The interest as a marker of endothelial dysfunction has increased in the last few years. In this paper, a method for serum ADMA quantification by liquid chromatography tandem mass spectrometry has been described. To test the utility in a pathological condition ADMA levels in hypertensive subjects have been measured.

Methods: HPLC separation was performed by hydrophilic interaction chromatography using acetonitrile/water containing 0.1% formic acid and 20 mmol/L ammonium formate. Selected reaction monitoring was performed following the transitions m/z 203.1→46.4 for ADMA and 210.1→46.3 for the internal standard [2H7]ADMA.

Results: The method was linear up to 10 μmol/L, limit of detection and limit of quantification were 0.005 μmol/L and 0.01 μmol/L, respectively. Recovery was higher than 96%. Intra- and inter-assay imprecision were lower than 6%. The accuracy, expressed as bias %, was <2.5. ADMA in "healthy" subjects ranged from 0.343 to 0.608 μmol/L and resulted significantly lower than that measured in hypertensive subjects (p<0.001).

Conclusions: The method developed is selective and sensitive, thus suitable not only for research purposes, but also for routinely work.

Download full-text PDF

Source
http://dx.doi.org/10.1515/CCLM.2011.691DOI Listing

Publication Analysis

Top Keywords

mass spectrometry
8
hypertensive subjects
8
endothelial dysfunction
8
adma levels
8
adma
5
determination asymmetric
4
asymmetric dimethyl
4
dimethyl arginine
4
arginine human
4
human serum
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!