Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Statement Of Problem: Clinicians must know if a new screw can predictably increase reverse torque after multiple screw insertion cycles.
Purpose: The purpose of this study was (1) to compare the effect of multiple implant prosthetic screw insertion and removal cycles on reverse torque, (2) to determine whether a new screw, after multiple screw insertion cycles, affects reverse torque, and (3) to assess implant and prosthetic screw thread surface morphology with scanning electron microscopy (SEM).
Material And Methods: One primary screw was paired with an implant (MT Osseospeed) and inserted to 25 Ncm torque 9, 19, 29, or 39 times (n=10). Primary screw reverse torque values were recorded after each insertion. A second, reference screw was then paired with each implant for a final screw insertion, and reverse torque was measured. Maximum, minimum, median, and mean values (P(max), P(min), P(median), and P(mean)) were identified for primary screws. A 1-way ANOVA and Tukey HSD post hoc analysis assessed the influence of multiple screw insertion cycles on P(max), P(min), P(median), and P(mean) values (α=.05). Confidence intervals were used to test differences between reference (REF) screw data and corresponding DMAX and DMIN (DMAX=P(max)-REF; DMIN=P(min)-REF). The surface topography of an unused implant and screw and of 1 implant and screw from each group was evaluated with SEM.
Results: Pairwise comparisons showed that 9 or fewer insertion cycles resulted in significantly greater mean reverse torque (20.9 ± 0.5 Ncm; P<.01). After 19, 29, or 39 cycles, the second, reference screw achieved significantly greater reverse torque than the minimum recorded values (P<.05). Implant thread surface morphology changes occurred primarily during the first 10 insertions.
Conclusions: After 10 screw insertion cycles, a new prosthetic screw should be used with the implant system tested to maximize screw reverse torque and maintain preload when an abutment is definitively placed.