Background: With increasing awareness for radiation exposure, the study of diagnostic accuracy of coronary CT angiography (CCTA) with low radiation dose techniques is mandatory to both radiologist and clinician.
Purpose: To compare diagnostic performance and effective radiation dose between step-and-shoot prospectively ECG-gated and retrospectively ECG-gated with tube current modulation (TCM) CCTA using 128-slice multidetector computed tomography (MDCT).
Material And Methods: We retrospectively evaluated 60 patients who underwent CCTA with either of two different low-dose techniques using 128-slice MDCT (23 patients for step-and shoot-prospectively ECG-gated and 37 patients for retrospectively ECG-gated with TCM CCTA) followed by conventional coronary angiography. All coronary arteries and all segments thereof, except anatomical variants or small size (< 1.5 mm) ones, were included in analysis.
Results: In per-segment analysis, sensitivity, specificity, positive predictive value, and negative predictive value were 91/96%, 95/94%, 75/73%, and 98/99% for step-and-shoot prospectively ECG-gated and retrospectively ECG gated with TCM CCTA, respectively, relative to conventional coronary angiography. Effective radiation dose were 1.75 ± 0.83 mSv, 4.91 ± 1.71 mSv in the step-and-shoot prospectively ECG-gated and retrospectively ECG-gated with TCM CCTA groups, respectively.
Conclusion: The two low-radiation dose CCTA techniques using 128-slice MDCT yields comparable diagnostic performance for coronary artery disease in symptomatic patients with low heart rates.