Objective: While patients with cardiac implantable electronic devices could benefit from magnetic resonance (MR) imaging, the presence of such devices has been designated as an absolute contraindication to MR. Although scanning algorithms are proposed for cardiac implantable electronic devices, their safety remains uncertain. To address this issue, the safety of serial cardiac MR scans was evaluated in patients with implantable cardioverter defibrillators (ICDs).

Methods: Three serial cardiac MR scans were prospectively performed at 1.5 T on 10 patients (9 men) of median age 56 years (range 51-68) with ICDs. ICD interrogation was performed before and after the MR scan and at a follow-up of median 370 days (range 274-723). Image quality was also assessed.

Results: In all patients MR scanning occurred without complications. There were no differences between pre- and post-MR pacing capture threshold, pacing lead or high voltage lead impedance, or battery voltage values. During follow-up there were no occurrences of ICD dysfunction. Although most patients had image artifacts, the studies were generally diagnostic regarding left ventricular function and wall motion. Delayed enhancement imaging was of good quality for inferior wall and inferolateral infarcts, but ICD artifacts often affected the imaging of anterior wall infarcts.

Conclusion: Serial MR scans at 1.5 T in patients with ICDs, when carefully performed in a monitored setting, have no adverse effects on either patient or device. When required, single or multiple MR scans at 1.5 T may therefore be considered for clinical diagnostic purposes in these patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3994993PMC
http://dx.doi.org/10.1136/heartjnl-2011-300153DOI Listing

Publication Analysis

Top Keywords

safety serial
8
patients
8
patients implantable
8
implantable cardioverter
8
cardioverter defibrillators
8
cardiac implantable
8
implantable electronic
8
electronic devices
8
serial cardiac
8
cardiac scans
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!