Objective: To evaluate a sensor-augmented insulin pump with a low glucose suspend (LGS) feature that automatically suspends basal insulin delivery for up to 2 h in response to sensor-detected hypoglycemia.

Research Design And Methods: The LGS feature of the Paradigm Veo insulin pump (Medtronic, Inc., Northridge, CA) was tested for 3 weeks in 31 adults with type 1 diabetes.

Results: There were 166 episodes of LGS: 66% of daytime LGS episodes were terminated within 10 min, and 20 episodes lasted the maximum 2 h. LGS use was associated with reduced nocturnal duration ≤2.2 mmol/L in those in the highest quartile of nocturnal hypoglycemia at baseline (median 46.2 vs. 1.8 min/day, P = 0.02 [LGS-OFF vs. LGS-ON]). Median sensor glucose was 3.9 mmol/L after 2-h LGS and 8.2 mmol/L at 2 h after basal restart.

Conclusions: Use of an insulin pump with LGS was associated with reduced nocturnal hypoglycemia in those at greatest risk and was well accepted by patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3161284PMC
http://dx.doi.org/10.2337/dc10-2411DOI Listing

Publication Analysis

Top Keywords

insulin pump
16
nocturnal hypoglycemia
12
hypoglycemia greatest
8
greatest risk
8
lgs feature
8
lgs associated
8
associated reduced
8
reduced nocturnal
8
lgs
7
insulin
6

Similar Publications

Unlabelled: The hybrid closed-loop (HCL) system, Medtronic MiniMed 770G, has been available for use by Japanese individuals with type 1 diabetes mellitus since 2021. The aim of this study was to evaluate the effect of its use on glycemic variability and quality of life (QOL) in this population. This multicenter, open-label, prospective observational study included 14 Japanese individuals with type 1 diabetes mellitus treated with MiniMed 640G.

View Article and Find Full Text PDF

Background: The emergency medicine (EM) pharmacist is an integrated part of the Emergency Department (ED) interdisciplinary team in many countries, including Malaysia. The presence of EM pharmacists in the ED has positively impacted patient outcomes. Data on EM pharmacists' interventions is scarce in the Asian region.

View Article and Find Full Text PDF

A woman in her 40s presented with severe post-bariatric hypoglycaemia that persisted despite nutritional therapy and pharmacological therapy with acarbose and subcutaneous octreotide with meals. The nutritional limitations were difficult to sustain, and she developed adverse effects to the pharmacological therapy, and hence, doses could not be increased. She was subsequently treated with subcutaneous octreotide via an insulin pump, with a continuous basal rate and additional bolus doses with meals.

View Article and Find Full Text PDF

Ethnic disparities in HbA1c and hypoglycemia among youth with type 1 diabetes: beyond access to technology, social deprivation and mean blood glucose.

BMJ Open Diabetes Res Care

January 2025

Diabetes and Endocrinology, Birmingham Women's and Children's NHS Foundation Trust, Birmingham, UK

Introduction: The UK national pediatric diabetes audit reports higher HbA1c for children and young people (CYP) with type 1 diabetes (T1D) of Black ethnicity compared with White counterparts. This is presumably related to higher mean blood glucose (MBG) due to lower socioeconomic status (SES) and less access to technology. We aimed to determine if HbA1c ethnic disparity persists after accounting for the above variables.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!