Introduction: Increased accuracy when using fiducial markers for prostate localisation is well documented. This project aimed to establish the improvement in accuracy when using gold markers for daily prostate localisation, to assess patient satisfaction and morbidity from the transrectal implantation of gold seed markers and establish staff attitudes towards the newly introduced processes.
Methods: Twenty patients with prostate cancer had three gold seeds implanted into the base, apex and central zone of the prostate transrectally using ultrasound guidance. Surveys were conducted to assess staff and patient satisfaction with the process of gold seed localisation. The gold markers were used to localise the prostate on a daily basis using megavoltage electronic portal imaging. Measurements were taken to establish the increase in accuracy when using gold fiducial markers compared with using the surrounding bony anatomy.
Results: Inter-fraction motion (1 standard deviation (SD)) of the fiducial markers was 2.20, 4.28 and 4.27 mm in the LR, SI and AP directions, respectively. Intra-fraction prostate motion (1 SD) was measured as 0.8 mm LR, 1.1 mm SI and 2.0 mm AP. The patient survey showed that the insertion and associated side effects were acceptable, with 5% of patients stating that the seed insertion was worse than the prostate biopsy, and 23.1% of patients experienced short duration (1-2 days) haematuria. The staff survey showed that daily online image guidance was achievable without affecting patient throughput. Thirty percent of treatment staff believed that performing online daily localisation did not add any extra time to a standard treatment, and the remaining 70% thought that the added time was minimal (2-4 min).
Conclusions: Gold fiducial markers are an accurate, reliable and tolerable method of daily prostate localisation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1754-9485.2011.02278.x | DOI Listing |
J Clin Med
January 2025
Department of Cardiovascular Surgery, Houston Methodist Hospital, 6550 Fannin St, Houston, TX 77030, USA.
: The aim of this study was to compare the outcomes of stiff wire-based 2D3D, 3D3D image fusion (IF), and non-image fusion techniques for simple zone 2 and zone 3 TEVAR cases in terms of radiation exposure, contrast dose, and fusion and projection accuracy. : A single-center retrospective observational study was conducted based on data gathered from patients who underwent TEVAR between 2016 and 2023 at our tertiary aortic referral center. Those who underwent Z2 and Z3 TEVAR during the indicated period were included.
View Article and Find Full Text PDFCancers (Basel)
January 2025
Department of Ophthalmology, University of Lübeck, University Medical Center Schleswig-Holstein, Campus Lübeck, 23562 Lübeck, Germany.
: Accurate target definition, treatment planning and delivery increases local tumor control for radiotherapy by minimizing collateral damage. To achieve this goal for uveal melanoma (UM), tantalum fiducial markers (TFMs) were previously introduced in proton and photon beam radiotherapy. However, TFMs cause pronounced scattering effects in imaging that make the delineation of small tumors difficult.
View Article and Find Full Text PDFJ Prosthodont
January 2025
Department of Prosthodontics, Jordan University of Science & Technology, Irbid, Jordan.
Purpose: To investigate the feasibility and accuracy (trueness and precision) of facial scanning and virtual patient representation (VPR).
Materials And Methods: One participant was recruited and informed consent was obtained. VPR was performed 30 times with a custom fabricated intraoral scan body (ISB).
J Appl Clin Med Phys
January 2025
Department of Radiation and Cellular Oncology, University of Chicago Medical Center, Chicago, Illinois, USA.
Background: Various methods exist to correct for intrafraction motion (IFM) of the prostate during radiotherapy. We sought to characterize setup corrections in our practice informed by the TrueBeam Advanced imaging package, and analyze factors associated with IFM.
Methods: 132 men received radiation therapy for prostate cancer with a volumetric modulated arc therapy technique.
Heliyon
January 2025
BRITElab, Harry Perkins Institute of Medical Research, QEII Medical Centre Nedlands and Centre for Medical Research, The University of Western Australia, Perth, Australia.
Breast-conserving surgery accompanied by adjuvant radiotherapy is the standard of care for patients with early-stage breast cancer. However, re-excision is reported in 20-30 % of cases, largely because of close or involved tumor margins in the specimen. Several intraoperative tumor margin assessment techniques have been proposed to overcome this issue, however, none have been widely adopted.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!