Background: To compare intraocular pressure (IOP) measurements obtained by the Icare ONE rebound tonometer (RTONE) and the Goldmann applanation tonometer (GAT) in healthy persons and glaucoma patients in a prospective study, and to investigate the influence of central corneal thickness (CCT).

Methods: Measurements on 126 right eyes were obtained by three equally skilled ophthalmologists with each of the above-mentioned tonometers. In addition, patients measured their own IOP with the RTONE (RTONE(p)). The means and standard deviation for all tonometers were compared. Agreement between the tonometers was calculated using the Bland-Altman method.

Results: A total of 95 (75.3%) patients were able to perform correct self-tonometry. Mean IOPs obtained were 17.1 ± 5.9 mmHg (RTONE performed by ophthalmologist: RTONE (o)), 17.3 ± 5.6 mmHg (RTONE(p)) and 16.5 ± 5.1 mmHg (GAT). Correlation analysis indicated a good correlation between IOP readings obtained using RTONE(o) and RTONE(p) (ρ = 0.916; p < 0.001) and RTONE(o) and GAT (ρ = 0.901; p < 0.001). Bland-Altman analysis revealed a mean difference (bias) between RTONE(o) and RTONE(p), between RTONE(o) and GAT, and between RTONE(p) and GAT of -0.2, 0.6, and 0.8 mmHg, respectively, with 95% limits of agreement of -5.0 to 4.5, -4.4 to 5.6, and -4.6 to 6.1 mmHg, respectively. The difference between RTONE(o) and GAT significantly increased with increasing CCT (ρ = 0.004), with a 10% increase in CCT resulting in a 1.8% increase in the difference.

Conclusions: Measurements obtained with the RTONE, either by an ophthalmologist or by the patient, showed an excellent correlation with those provided by applanation tonometry. RTONE generally tends to overestimate IOP compared to GAT readings and displays a dependence on CCT. This study was registered with the DRKS (German Clinical Trials Register; www.germanctr.de ; DRKS00000478).

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00417-011-1785-7DOI Listing

Publication Analysis

Top Keywords

rebound tonometer
8
intraocular pressure
8
tonometer monitoring
4
monitoring intraocular
4
pressure background
4
background compare
4
compare intraocular
4
pressure iop
4
iop measurements
4
measurements icare
4

Similar Publications

Patient Perspectives on Home Tonometer Usage for Glaucoma.

Clin Ophthalmol

January 2025

Department of Ophthalmology and Visual Sciences, W. K. Kellogg Eye Center, University of Michigan Medical School, Ann Arbor, MI, USA.

Purpose: To understand patient experiences using the iCare HOME tonometer through assessing ease of use, device usage patterns, proficiency, and patient comments.

Methods: We conducted a prospective survey-based cohort study of 19 patients (35 eyes) with glaucoma or glaucoma suspect diagnoses. Patients received training on home tonometer usage and were loaned the device for one week to measure intraocular pressure at home multiple times daily.

View Article and Find Full Text PDF

Aims/purpose: This study aims to assess whether changes in intraocular pressure (IOP) occur after a blood draw and quantify these changes' magnitude in a group of patients over 60 years old.

Methods: We analysed data from the right eye of 54 subjects, with a mean ( SD) age of 68 4 years. The measurements were conducted in the morning while fasting.

View Article and Find Full Text PDF

Background: Bacterial contamination of medical equipment is a significant risk in both human and veterinary medicine, leading to potential cross-contamination between clinicians, technicians, patients, and healthcare professionals. Pathogens can persist on surfaces, resulting in healthcare-associated infections. Tonometry used to measure intraocular pressure for diagnosing conditions like glaucoma and uveitis, is crucial in both human and veterinary care.

View Article and Find Full Text PDF

Purpose: This study aims to perform a clinical investigation of an innovative rebound technology-based device, the M-TONX, to simultaneously measure intraocular pressure (IOP) and central corneal thickness (CCT).

Methods: The IOP and CCT of the patients were first measured by the M-TONX. Then, the measurements were repeated by the Goldman applanation (GAT) and the Pentacam corneal topographer, as the standard devices.

View Article and Find Full Text PDF

Purpose: To investigate the agreement between the Icare HOME tonometer and the Goldmann Applanation Tonometry (GAT) in assessing the Peak Intraocular Pressure during the Water-Drinking Test (WDT).

Design: Prospective cross-sectional study.

Subjects: 79 eyes of 41 open-angle glaucoma patients were included in the study.

View Article and Find Full Text PDF