Development of a point-of-care assay system for measurement of presepsin (sCD14-ST).

Clin Chim Acta

Research and Development Division, Mitsubishi Chemical Medience Corporation, Japan.

Published: November 2011

AI Article Synopsis

Article Abstract

Background: The soluble CD14 subtype (sCD14-ST: renamed as presepsin) is a novel soluble CD14 molecule that is useful for diagnosing sepsis because sCD14-ST levels increase specifically in sepsis patients.

Methods: A fully automated PATHFAST® Presepsin assay system based on a chemiluminescent enzyme immunoassay was developed for detecting presepsin in human whole blood.

Results: The limit of blank, limit of detection, and limit of quantification were 2.33, 13.4, and 47.6 pg/ml, respectively. The assay linearity was achieved up to 20,000 pg/ml. Intra-assay imprecision was 3.4-4.8% for plasma and 2.7-7.1% for whole blood. Within-run imprecision and total imprecision for plasma were 3.6-4.4% and 5.2-6.5%, respectively. No interference was observed with bilirubin, hemoglobin, lipids, triglyceride, or rheumatoid factors. The reference intervals (95% percentile, n=127) were 333 pg/ml for plasma and 314 pg/ml for whole blood. The PATHFAST® Presepsin assay correlated well with a previously reported two-step presepsin ELISA (r=0.984, n=40). Furthermore, the concentration of presepsin was significantly higher in the sepsis group than in the healthy group.

Conclusion: The PATHFAST® Presepsin assay performed well and can be used for point-of-care.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2011.07.024DOI Listing

Publication Analysis

Top Keywords

pathfast® presepsin
12
presepsin assay
12
assay system
8
presepsin
8
soluble cd14
8
assay
5
development point-of-care
4
point-of-care assay
4
system measurement
4
measurement presepsin
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!