A newly developed recombinant antigen-based anti-HIV-1/HIV-2 enzyme immunoassay (Abbott Recombinant HIV-1/HIV-2 EIA) was evaluated against a second generation anti-HIV-1 EIA (Abbott Recombinant HIV-1 EIA). Five thousand and twenty-nine sera from European blood donors and 403 sera from central African blood donors were used in the evaluation, along with four panels and one cohort. The panels included 99 'problem' sera, 733 sera with antibodies to HIV-1 from asymptomatic people and from patients at different disease stages, 25 serial bleeds from five plasmapheresis donors seroconverting for antibodies to HIV-1, and 202 sera with antibodies to HIV-2 collected from healthy and diseased people of European or west African origin. In addition, 734 sera collected from a west African cohort were tested. Using Western blot as the reference standard, the specificity obtained by the recombinant anti-HIV-1 EIA (HIV-i EIA) was 99.90% [99.81-99.99%; 95% confidence limits (95% CL)] with European blood donor sera; 99.50% (98.78-100%) with Central Africa blood donor sera; 92.93% (87.78-98.08%) with 'problem' sera and 99.43% (98.87-100%) with sera from a west African cohort. Using the same samples, the recombinant anti-HIV-1/HIV-2 EIA (HIV-1/HIV-2 EIA) yielded a specificity of 99.84% (99.73-99.95%), 99.50% (98.78-100%), 95.96% (92.00-99.92%) and 98.58% (97.69-99.47%), respectively. All 776 Western blot-confirmed anti-HIV-1 sera were reactive in both EIAs, and the EIA-reactive samples from seroconverting plasma donors were always observed for both assays in the same serial bleed. For HIV-2, the HIV-1 EIA yielded an overall sensitivity of 75.83% (69.93-81.72%) compared with 99.53% (98.58-100%) for HIV-1/HIV-2 EIA. The addition of a recombinant env-protein of HIV-2 to the recombinant env and core proteins of HIV-1 on the solid phase of HIV-1 EIA improved the detection of anti-HIV-2 while preserving the assay's overall specificity and sensitivity for the detection of anti-HIV-1.

Download full-text PDF

Source
http://dx.doi.org/10.1097/00002030-199002000-00006DOI Listing

Publication Analysis

Top Keywords

hiv-1/hiv-2 eia
12
hiv-1 eia
12
west african
12
sera
11
eia
10
recombinant
8
enzyme immunoassay
8
abbott recombinant
8
anti-hiv-1 eia
8
european blood
8

Similar Publications

Background: Sexually transmitted infections (STIs) such as syphilis and HIV remain to be a significant public health issue worldwide. Dual rapid point-of-care tests (POCTs) have shown promise for detecting antibodies to HIV and syphilis but have not been fully evaluated in the field. Our study supported the WHO ProSPeRo study on Sexually Transmitted Infection Point-of-Care Testing (STI POCT) by providing external quality assessment (EQA) for HIV and syphilis testing in reference laboratories and their associated clinical sites in seven countries.

View Article and Find Full Text PDF

Background: A single, simplified approach for human immunodeficiency virus (HIV)-1/HIV-2 antibody confirmation/differentiation is needed for the HIV blood donation supplemental algorithm used in the United States. A clinical evaluation of the Geenius assay was performed-the same assay used for HIV diagnostic confirmation/differentiation in the United States since 2014.

Study Design And Methods: Well-characterized unlinked donation samples classified as HIV negative, false positive, or confirmed positive were included in the study: 200 antibody-nonreactive, 200 HIV-1 immunofluorescence assay (IFA) confirmed-positive, and 100 antibody-screen false-positive donations, equally divided between serum and plasma.

View Article and Find Full Text PDF

Laboratory assays that can accurately distinguish recent (occurring within the past year) from long-standing (>1 year) HIV infection are crucial for understanding HIV transmission dynamics in a population. However, often these efforts are confounded by inaccurate HIV diagnosis and the presence of HIV-2 in the population being surveyed. This study describes development of a multiplex assay that can simultaneously perform HIV diagnosis, HIV serotyping, and detection of recent HIV-1 infection in a single well.

View Article and Find Full Text PDF

There has been increasing recognition of the importance of diagnosing individuals during the earliest stages of human immunodeficiency virus (HIV) infection. Sera from individuals referred to a primary HIV infection research program were screened using the IgG-sensitive Vironostika HIV-1 Microelisa System, IgG/IgM-sensitive GS HIV-1/HIV-2 Plus O antibody enzyme immunoassay (EIA), or Abbott ARCHITECT HIV antigen (Ag)/antibody (Ab) Combo assay and confirmed by the Bio-Rad Multispot and Western blot. A subset of participants was co-enrolled in a study designed to compare the ability of point-of-care tests to detect early infection.

View Article and Find Full Text PDF

Heterophilic interference in specimens yielding false-reactive results on the Abbott 4th generation ARCHITECT HIV Ag/Ab Combo assay.

J Clin Virol

July 2018

National Laboratory for HIV Reference Services, National Microbiology Laboratory, Winnipeg, MB, Canada. Electronic address:

Background: False-reactivity in HIV-negative specimens has been detected in HIV fourth-generation antigen/antibody or 'combo' assays which are able to detect both anti-HIV-1/HIV-2 antibodies and HIV-1 antigen.

Objectives: We sought to characterize these specimens and determine the effect of heterophilic interference.

Study Design: Specimens previously testing as false-reactive on the Abbott ARCHITECT HIV Ag/Ab combo assay and re-tested on a different (Siemens ADVIA Centaur HIV Ag/Ab) assay.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!