Finite element analysis is a useful analytical tool for the design of biomedical implants. The aim of this study was to investigate the behavior of temporomandibular joint implants with multiple design variables of the screws used for fixation of the implant. A commercially available implant with full mandible was analyzed using a finite element software package. The effects of different design variables such as orientation, diameter and stem length of the screws on the stress distribution in bone for two different surgical procedures were investigated. Considering the microstrain in bone as a principal factor, the acceptable ranges for screw diameter and length were determined. Parallel orientation of the screws performed better from a stress point of view when compared to the zig-zag orientation. Sufficient contact between the implant collar and mandibular condyle was shown to reduce the peak stresses which may lead to long term success. The distance between screw holes in the parallel orientation was much closer when compared to the zig-zag orientation. However, the stresses in bone near the screw hole area for the parallel orientation were within acceptable limits.