AI Article Synopsis

  • Five cases of HIV-1 RNA-positive blood donations were undetected by three different nucleic acid amplification tests (NATs), leading to two HIV-1 transmissions to recipients.
  • Investigations revealed that the false-negative results were due to mismatches between the viral variants and the NAT assay primers and probes used.
  • The study suggests that dual-target NAT assays are more effective at detecting HIV-1 variants compared to monotarget assays, prompting consideration for their use in blood donation screening in Germany.

Article Abstract

Background: Five cases of human immunodeficiency virus Type 1 (HIV-1) RNA-positive blood donations are described that escaped detection by three different CE-marked nucleic acid amplification technique (NAT) screening assays. These events were associated with two HIV-1 transmissions to recipients of blood components. The implicated NAT assays are monotarget assays and amplify in different viral genome regions (group-specific antigen or long terminal repeat). Investigations into the cause of the false-negative test results were initiated.

Study Design And Methods: Plasma specimens of the five NAT false-negative cases were comparatively investigated in 12 CE-marked HIV-1 NAT systems of differing design. The relative amplification efficiency for the HIV-1 variant was determined for each assay. Sequencing of the variants in the region targeted by each false-negative NAT assay allowed comparison with the respective primers and probes.

Results: Some of the NAT assays designed in a similar way to false-negative monotarget NATs also revealed deficiencies in detecting the viral variants. In each case sequencing of the assay target region in the variants demonstrated mismatches with primers and probes used by the assays. Some dual-target assays showed decreased amplification efficiency, but not false-negative results.

Conclusion: HIV is characterized by its rapid evolution of new viral variants. The evolution of new sequences is unpredictable; NAT screening assays with a single target region appear to be more vulnerable to sequence variations than dual-target assays. Based on this experience with false-negative tests results by monotarget NAT assays, the Paul-Ehrlich-Institut is considering requesting dual-target NAT assays for HIV-1 blood donation screening in Germany.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1537-2995.2011.03281.xDOI Listing

Publication Analysis

Top Keywords

nat assays
16
assays
10
nat
9
nucleic acid
8
acid amplification
8
human immunodeficiency
8
immunodeficiency virus
8
virus type
8
nat screening
8
screening assays
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!