Objectives: The purpose was to assess thermal changes and drill wear in bovine bone tissue with the use of twisted stainless steel and zirconia-based drills, during implant site preparation.
Methods: A total of 100 implant site preparations were performed on bovine ribs using a surgical unit linked to a testing device, in order to standardize/simulate implant drilling procedures. Bone temperature variations and drilling force were recorded when drilling at a depth of 8 and 10 mm. A constant irrigation of 50 ml/min. (21±1°C) and drilling speed of 800 r.p.m. were used. Scanning electron microscopy analysis was preformed prior and after drilling.
Results: Mean temperature increase with both drills at 8 mm was 0.9°C and at 10 mm was 2°C (P<0.0001). Statistical significant higher bone temperatures were obtained with stainless steel drill (1.6°C), when comparing with the ceramic drill (1.3°C) (P<0.05). Temperature increase was correlated with higher number of perforations (P<0.05) and drilling load applied. There was no significant association between drilling force applied and temperature increase by either drill or at either depth. No severe signs of wear of either drill were detected after 50 uses.
Conclusions: Drill material and design, number of uses, depth and drilling load applied appear to influence bone temperature variations during implant site preparation. Drilling depth was a predominant factor in bone temperature increase. Both drills can be used up to 50 times without producing harmful temperatures to bone tissue or severe signs of wear and deformation.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1600-0501.2011.02248.x | DOI Listing |
Cureus
November 2024
Cardiology, Akita Cerebrospinal and Cardiovascular Center, Akita, JPN.
Background Ventricular septal pacing has long been performed using a stylet during pacemaker implantation, but with the availability of guiding catheters, His bundle pacing and left bundle branch area pacing have also been performed. However, it is not known to what extent the tip load of the ventricular lead differs when a guiding catheter is used compared with a stylet alone. In this study, the tip load was measured for different stylet stiffness and guiding catheter geometries at sites where His bundle pacing and left bundle branch area pacing were assumed.
View Article and Find Full Text PDFCureus
November 2024
Trauma and Orthopaedics, North Manchester General Hospital, Manchester, GBR.
Introduction: Salvage arthroplasty for failed proximal femoral fracture fixation is a complex procedure. This involves the removal of the primary failed or broken implant followed by a hip joint replacement procedure. The complications and technical difficulties associated with these surgeries are often difficult to anticipate.
View Article and Find Full Text PDFCureus
November 2024
Anesthesiology, Anesthesia Unit, National Cancer Institute, Santiago, CHL.
Background: Totally implantable venous access devices (TIVADs) are widely used in oncology patients to facilitate central venous access. Although they offer benefits, TIVADs can be associated with complications.
Materials And Methods: This retrospective cohort study included all oncology patients 18 years or older who underwent TIVAD implantation between September 2015 and October 2019.
Sci Rep
December 2024
Student Research Committee, Department of Epidemiology, School of Public Health and Safety, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Very high heat is generated during the polymerization of poly (methyl methacrylate) (PMMA) bone cement, which is used for implant fixation in orthopedic surgery. As such, it has been suggested that irrigating the bone cement layer in the surgical site with a saline solution is a way of cooling the layer. In this study, we aimed to determine the influence of irrigation with a saline solution on the flexural strength and the microstructure of the test specimens of two PMMA bone cement brands: Simplex P and FIX 1.
View Article and Find Full Text PDFInt J Biol Macromol
December 2024
Department of Animal Genetics and Breeding, National Engineering Laboratory for Animal Breeding, College of Animal Science and Technology, China Agricultural University, Beijing 100193, China; Key Laboratory of Animal Genetics, Breeding and Reproduction of the Ministry of Agriculture and Rural Affairs, College of Animal Science and Technology, China Agricultural University, Beijing 100193, China; CAU-SC Advanced Agricultural & Industrial institute, CAU-SCCD Advanced Agricultural & Industrial institute, China Agricultural University, Chengdu 611430, China. Electronic address:
Litter size in pigs is affected by factors such as ovulation number, embryonic survival, and uterine environment conditions. Endometrial epithelial and stromal cells represent the first site of contact between the embryo and sows; therefore, dynamic changes in the growth and development of these cells are among the major factors affecting the intrauterine environment and implantation. Bone morphogenetic protein receptor type-1B (BMPR1B) is a receptor of the bone morphogenetic protein (BMP) family that has been identified as a candidate gene for reproductive traits in pigs.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!