Background: In 2007, Medtronic Sprint Fidelis defibrillator leads were taken off the market due to a high rate of lead failure. Current data do not allow for risk stratification of patients with regard to lead failure.
Aims: We sought to determine predictors of Sprint Fidelis lead failure.
Methods: Between 2004 and 2007, 269 Sprint Fidelis leads were implanted in 258 patients in our centre. Variables associated with lead failure were assessed by the Kaplan-Meier method and a Cox survival model.
Results: During a median follow-up of 2.80 years (maximum 5.32), we observed 33 (12.3%) Sprint Fidelis lead failures (5-year survival, 65.6% ± 7.5%). In univariate analysis, age was the only predictor of lead failure (hazard ratio [HR] for 1-year increase 0.97; 95% confidence interval [CI] 0.95-0.99; p=0.009). Patients aged<62.5 years (median) had a significantly increased risk of lead failure compared with patients aged>62.5 years (HR 2.80; CI 1.30-6.02; p=0.009). Survival without Sprint Fidelis lead failure was 55.6% ± 10.4%) in patients aged<62.5 years (24/134 leads) vs 78.6% ± 8.8% in patients aged>62.5 years (9/135 leads). The annual incidence of lead failure in patients aged<62.5 years was 11.6% ± 4.9% during the fourth year after implantation and 22.9% ± 13.2% during the fifth year.
Conclusion: Overall, we found a higher rate of Sprint Fidelis lead dysfunction than previously described. Lead failure was much more frequent in younger patients. Our results emphasize the need for close follow-up of younger patients with Sprint Fidelis leads and suggest that, in these patients, the implantation of a new implantable cardioverter defibrillator lead at the time of generator replacement might be reasonable.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.acvd.2011.05.002 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!