Background: The newly developed balloon-expandable Mguard stent system, a combination of an ultra-thin polymer mesh sleeve attached to the external surface of a BMS, was conceived to provide embolic protection during PCI of SVG and thrombus-containing lesions. Although the acute results (<30 days) have pointed to the efficacy of this novel device, few is known about its long-term performance.

Methods: The present article address the 1-year clinical results of a cohort of 30 patients enrolled in the INSPIRE trial. Inclusion critiria was de novo lesions in SVG or native vessels with angiographic evidence of instability with potential to provoke flow disturbances and/or distal embolization. The primary endpoint (incidence of MACE-composite of cardiac death, nonfatal MI, and TLR) up to 30 days of the procedure has already been published. Secondary endpoints here presented included in-stent late lumen loss (QCA), % of stent obstruction (IVUS) at 6 months and combined MACE at 1 year. QCA and IVUS were performed by independent corelabs.

Results: Mean population age was 63 years with 38% of diabetics. Overall, 55% presented with ACS and 57% of lesions were located in SVG. Most lesions had complex morphology including the presence of thrombus (26%) and ulceration (20%). Distal/proximal protection devices were not used. Preprocedural QCA data showed lesion length and reference vessel diameter of 12.0 ± 4.5 mm and 3.0 ± 0.5 mm. The MGuard stent was successfully delivered in all cases and final TIMI-3 was achieved in 100% with no MACE up to 30 days. At 6 months, in-stent late loss and % of stent obstruction were 1.0 ± 0.4 mm and 28.5 ± 15.6%. Up to 1 year there was no case of cardiac death, two MI (one Q-wave and one non-Q-wave) and six cases of ischemia-driven TLR. Of note, there was no case of definite/probable stent thorombosis.

Conclusions: In this series of patients treated with MGuard stent, the novel device showed no midterm efficacy and safety concerns.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.22980DOI Listing

Publication Analysis

Top Keywords

mguard stent
8
one-year inspire
4
inspire trial
4
trial novel
4
novel mguard
4
stent serial
4
serial analysis
4
analysis qca
4
qca ivus
4
ivus background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!