Purpose: To test megavolt cone-beam CT (MV-CBCT) in order to evaluate setup errors in prostate radiotherapy.

Patients And Methods: The setup of 9 patients was verified weekly by electronic portal imaging (EPI) and MV-CBCT, both per-formed in the same treatment session. EPI were compared with digitally reconstructed radiographies (DRRs). MV-CBCTs were matched to simulation CTs by manual registration based on bone markers (BMR), by manual registration based on soft tissues (STR) - rectum, bladder, and seminal vesicles - and by automatic registration (AR) performed by a mutual information algorithm. Shifts were evaluated along the three main axes: anteroposterior (AP), craniocaudal (CC), and laterolateral (LL). Finally, in 4 additional patients showing intraprostatic calcifications, the calcification mismatch error was used to evaluate the three MV-CBCT matching methods.

Results: A total of 50 pairs of orthogonal EPIs and 50 MV-CBCTs were analyzed. Assuming an overall tolerance of 2 mm, no significant differences were observed comparing EPI vs BMR in any axis. A significant difference (p < 0.001) was observed along the AP axis comparing EPI vs AR and EPI vs STR. On the calcification data set (22 measures), the calcification mismatch along the AP direction was significantly lower (p < 0.05) after STR than after BMR or AR.

Conclusion: Bone markers were not an effective surrogate of the target position and significant differences were observed comparing EPI or BMR vs STR, supporting the assessment of soft tissue position by MVCBs to verify and correct patient setup in prostate radiotherapy.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00066-011-2241-7DOI Listing

Publication Analysis

Top Keywords

comparing epi
12
megavolt cone-beam
8
manual registration
8
registration based
8
bone markers
8
calcification mismatch
8
differences observed
8
observed comparing
8
epi bmr
8
epi
6

Similar Publications

Background: The estimation of glomerular filtration rate (eGFR) is essential in the early detection of diabetic nephropathy. We herein compare the performance of common eGFR formulas against a gold standard measurement of GFR in patients with diabetes mellitus.

Methods: GFR was measured in 93 patients with diabetes mellitus using iohexol clearance as the reference standard.

View Article and Find Full Text PDF

Background: Proenkephalin A 119-159 (penKid) is a novel blood biomarker for real-time assessment of kidney function and was found to be independently associated with worsening kidney function and mortality. A novel penKid-based estimated glomerular filtration rate equation (eGFR), outperforms current creatinine-based eGFR equations in predicting iohexol or iothalamate plasma clearance-based measured GFR. In this study, we aimed to evaluate the predictive value of penKid and eGFR for all-cause mortality in stable patients at high cardiovascular risk.

View Article and Find Full Text PDF

Value of Myocardial Strain in Monitoring Fluorouracil-Based Chemotherapy-Related Cardiac Dysfunction in Gastrointestinal Cancer Patients.

Chin Med Sci J

November 2024

State Key Laboratory for Innovation and Transformation of Luobing Theory; Key Laboratory of Cardiovascular Remodeling and Function Research, Chinese Ministry of Education, Chinese National Health Commission and Chinese Academy of Medical Sciences; Department of Cardiology, Qilu Hospital of Shandong University, Jinan 250012, China.

Objectives: To investigate the predictive value of myocardial strain for cardiotoxicity associated with fluorouracil-based chemotherapies in gastrointestinal cancer patients.

Methods: Patients with diagnosis of gastrointestinal cancers, who were hospitalized for chemotherapy involving antimetabolic drugs, were eligible in this prospective study. Echocardiography was performed before and after each chemotherapy cycle during hospitalization until the completion of chemotherapy.

View Article and Find Full Text PDF

Background: Diffusion-weighted (DW) turbo-spin-echo (TSE) imaging offers improved geometric fidelity compared to single-shot echo-planar-imaging (EPI). However, it suffers from low signal-to-noise ratio (SNR) and prolonged acquisition times, thereby restricting its applications in diagnosis and MRI-guided radiotherapy (MRgRT).

Purpose: To develop a joint k-b space reconstruction algorithm for concurrent reconstruction of DW-TSE images and the apparent diffusion coefficient (ADC) map with enhanced image quality and more accurate quantitative measurements.

View Article and Find Full Text PDF

Background: Dysfunction of the glymphatic system (GS), a recently discovered brain by-product elimination system, is considered to be one of the pathophysiological mechanisms for common neurodegenerative diseases such as Alzheimer's disease (AD), dementia with Lewy bodies (DLB) and Parkinson's disease (PD). In 2017 a new way to assess the GS was proposed - a diffusion tensor images analysis along perivascular spaces (DTI-ALPS). In our work we evaluated the DTI-ALPS index in groups of patients with AD, DLB, PD and in a comparison group of patients with normal pressure hydrocephalus (NPH).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!