Purpose: Three dimensional dose polymer gel dosimetry measurements provide unique information on sophisticated dose distributions. In this study, the authors propose a novel method to improve the accuracy of polymer gel dosimeters by inserting a plastic scintillation detector (PSD) to provide a dose reference.

Methods: PSD dosimeters were calibrated using chromatic deconvolution and then inserted into polyacrylanide gel (PAG) dosimeters. The gel and the PSDs were immersed into water and irradiated with 6 MV wedge filtered beams to obtain a wide range of dose variation. Calibration vials containing the same gel were also irradiated to generate a standard calibration curve. The distribution of magnetic nuclear transverse relaxation rate (R2) values of the gel was determined with a multislice multiecho MRI sequence at 1.5 T. Another calibration curve was obtained by assigning the R2 values in the gel surrounding the scintillators to the dose determined by the PSDs. A reference calibration point from a PSD located in a low dose gradient area served to correct the standard calibration method yielding three novel calibration methods. The results were compared with EBT2 GAFCHROMIC film measurements acquired in the same condition and with the Pinnacle3 treatment planning dose calculations,

Results: The mean absolute error of the standard calibration method ranged from 6.1 to 12.4%. The corresponding gamma index (3%/3 mm distance to agreement) criterion was satisfied for only 56% of the pixels in the middle slice of the gel compared to Pinnacle3 dose calculations and to EBT2 film measurements in the center part of the field. Calibration methods using a PSD reduced the mean absolute error to less than 4%; this value was under 2.6% for one of the three methods. In that case, 98% of the pixels satisfied the gamma index criterion.

Conclusions: The accuracy of PAG dosimeters may be highly improved using one reference dose point measurement using a plastic scintillation detector. The best calibration procedure corrected the slope of the calibration curve derived from the calibration vials to match the R2 value around a PSD calibration, while keeping the R2 value at 0 Gy constant.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.3578601DOI Listing

Publication Analysis

Top Keywords

polymer gel
12
plastic scintillation
12
scintillation detector
12
calibration
12
standard calibration
12
calibration curve
12
gel
9
dose
9
pag dosimeters
8
calibration vials
8

Similar Publications

Evaporation-Assisted Synthesis of Olympic Gels.

Angew Chem Int Ed Engl

January 2025

UESTC: University of Electronic Science and Technology of China, Institute of Fundamental and Frontier Sciences, Jianshe Road, Chengdu, CHINA.

Catenated networks exclusively composed of intertwining rings were first envisioned as "Olympic gels" by Pierre-Gilles de Gennes four decades ago but have not been successfully prepared in artificial materials yet due to the challenge in synthesis. Herein, we present a bio-inspired, evaporation-assisted strategy to address this issue. In our design, the evaporation of liquid catalysts that induce ring-chain equilibrium of polymer systems drives macrocycles to encounter and assists their catenation through reversible cyclization.

View Article and Find Full Text PDF

Background: Spinal cord injury (SCI) treatment remains a formidable challenge, as current therapeutic approaches provide only marginal relief and fail to reverse the underlying tissue damage. This study aims to develop a novel composite material combining enzymatic nanoparticles and nerve growth factor (NGF) to modulate the immune microenvironment and enhance SCI repair.

Methods: CeMn nanoparticles (NP) and CeMn NP-polyethylene glycol (PEG) nanozymes were synthesized via sol-gel reaction and DSPE-mPEG modification.

View Article and Find Full Text PDF

Background: Disinfection of the root canal system is a challenge to all clinicians, calcium hydroxide Ca(OH) one of the most popular intracanal medications used for this purpose, has some unwanted effects on dentine. This study aimed to investigate the antibiofilm effect of Nanochitosan (CSNPs) and Calcium hydroxide Ca(OH) intra canal medications and their effect on the microhardness and chemical structure of radicular dentine.

Methodology: A total of 52 extracted human mandibular premolars were used.

View Article and Find Full Text PDF

Fast 3D printing of fine, continuous, and soft fibers via embedded solvent exchange.

Nat Commun

January 2025

Department of Mechanical Science and Engineering, The Grainger College of Engineering, University of Illinois Urbana-Champaign, Urbana, IL, USA.

Nature uses fibrous structures for sensing and structural functions as observed in hairs, whiskers, stereocilia, spider silks, and hagfish slime thread skeins. Here, we demonstrate multi-nozzle printing of 3D hair arrays having freeform trajectories at a very high rate, with fiber diameters as fine as 1.5 µm, continuous lengths reaching tens of centimeters, and a wide range of materials with elastic moduli from 5 MPa to 3500 MPa.

View Article and Find Full Text PDF

Corrigendum to "Facile degradation of chitosan-sodium alginate-chromium (III) gel in relation to leather re-tanning and filling" [International Journal of Biological Macromolecules, 240 (2023) 124437].

Int J Biol Macromol

January 2025

State Key Laboratory of Fluid Power & Mechatronic System, Key Laboratory of Soft Machines and Smart Devices of Zhejiang Province, Center for X-Mechanics, Department of Engineering Mechanics, Zhejiang University, Hangzhou 310027, PR China. Electronic address:

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!