Under analysis were causes of reoperations in bradyarrhythmic forms of disturbances of conductivity of the heart. 565 operations connected with electrocardiostimulation were performed on 276 patients. Reoperations in patients with the myocardial stimulation were made on 63 patients, with the endocardial stimulation--in 226 patients. Causes of reoperations were as follows: depletion of the source of ECS--31.5%; dislocation of the electrode--8.2%; damage of the electrode--15%, suppuration of the bed, decubitus of the stimulator--28.8%; competing rate--7.4%; break of the contact in the connecting system of the electrode and cardiostimulator--9.4%.