Objective: The purpose of this study was to investigate the incidence of mechanical complications associated with low-profile subcutaneous implantable venous access devices in gynecologic oncology patients.

Methods: Gynecologic oncology patients with low-profile Port-a-Caths implanted between March 2005 and July 2006 were identified into a computerized database. Patient demographics, operative complications, number of chemotherapy cycles, duration of implantation, and mechanical complications were collected. Primary outcomes included port leakage, catheter fracture, and catheter embolization.

Results: 112 patients underwent 115 Port-a-Cath placements with low profile single-lumen plastic ports with Groshong-valved catheters. Mean Port-a-Cath indwelling duration was 197 days (range: 4-395) with a mean number of 12 chemotherapy cycles (range 0-64). The cumulative complication rate necessitating removal or replacement was 15%. Of the 14 Port-a-Caths removed, ten (8.7%) were secondary to mechanical malfunction: one for leakage at the port site, two for catheter fracture, and seven for fracture with catheter embolization to the heart or pulmonary vasculature-most commonly the right ventricle. Patients with embolization were asymptomatic and all embolized catheters were successfully retrieved by interventional radiology without complications.

Conclusions: The rates of catheter fracture and embolization have previously been reported to be low in patients with subcutaneous Port-a-Caths, and have not been studied in patients receiving low-profile subcutaneous Port-a-Caths. This study suggests that catheter fracture may be more common (8.7%) and must be considered in patients with malfunctioning low-profile Port-a-Caths. Embolized catheters can be removed by interventional radiology without significant adverse affects.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ygyno.2011.06.012DOI Listing

Publication Analysis

Top Keywords

catheter fracture
16
low-profile subcutaneous
12
incidence mechanical
8
mechanical malfunction
8
subcutaneous implantable
8
implantable venous
8
venous access
8
access devices
8
patients receiving
8
mechanical complications
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!