A commercial 2D array seven29 detector has been characterized and its performance has been evaluated. 2D array ionization chamber equipped with 729 ionization chambers uniformly arranged in a 27 × 27 matrix with an active area of 27 × 27 cm² was used for the study. An octagon-shaped phantom (Octavius Phantom) with a central cavity is used to insert the 2D ion chamber array. All measurements were done with a linear accelerator. The detector dose linearity, reproducibility, output factors, dose rate, source to surface distance (SSD), and directional dependency has been studied. The performance of the 2D array, when measuring clinical dose maps, was also investigated. For pretreatment quality assurance, 10 different RapidArc plans conforming to the clinical standards were selected. The 2D array demonstrates an excellent short-term output reproducibility. The long-term reproducibility was found to be within ±1% over a period of 5 months. Output factor measurements for the central chamber of the array showed no considerable deviation from ion chamber measurements. We found that the 2D array exhibits directional dependency for static fields. Measurement of beam profiles and wedge-modulated fields with the 2D array matched very well with the ion chamber measurements in the water phantom. The study shows that 2D array seven29 is a reliable and accurate dosimeter and a useful tool for quality assurance. The combination of the 2D array with the Octavius phantom proved to be a fast and reliable method for pretreatment verification of rotational treatments.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.meddos.2010.12.013 | DOI Listing |
J Med Phys
January 2019
Department of Radiation Oncology, University of Texas Health San Antonio, San Antonio, TX, USA.
Introduction: Clinical evaluation of a two-dimensional (2D) liquid-filled ion chamber detector array used in the verification of highly modulated small beams of stereotactic body radiation therapy (SBRT) has been conducted.
Materials And Methods: Measurements with the Octavius 1000 SRS (PTW, Freiburg, Germany) detector with 977 liquid-filled ion chambers were compared against EDR2 film and PTW Octavius Seven29. The performance of detector array has been evaluated on ten SBRT patient plans.
Br J Radiol
October 2019
Department of Biomedical Engineering, Amirkabir University of Technology, Tehran, Iran.
Objective: The accuracy of dose delivery for intensity modulated radiotherapy (IMRT) treatments should be determined by an accurate quality assurance procedure. In this work, we used artificial neural networks (ANNs) as an application for the pre-treatment dose verification of IMRT fields based two-dimensional-fluence maps acquired by an electronic portal imaging device (EPID).
Methods: The ANN must be trained and validated before use for the pretreatment dose verification.
J Med Phys
January 2017
Department of Medical Physic Radiology and Physical Medicine, Faculty of Medicine/Dentistry, University of Murcia, Murcia, Spain.
Introduction: The aim of this work is to verify the use of radiochromic film in the quality assurance (QA) of volumetric-modulated arc therapy (VMAT) lung stereotactic body radiation therapy (SBRT) plans and compare the results with those obtained using an ion chamber array.
Materials And Methods: QA was performed for 14 plans using a two-dimensional-array seven29 and EBT3 film. Dose values per session ranged between 7.
Br J Radiol
May 2016
3 Department of Medical Physics, Royal Surrey County Hospital NHS Foundation Trust, Guildford, UK.
Objective: To demonstrate the benefit of complexity metrics such as the modulation complexity score (MCS) and monitor units (MUs) in multi-institutional audits of volumetric-modulated arc therapy (VMAT) delivery.
Methods: 39 VMAT treatment plans were analysed using MCS and MU. A virtual phantom planning exercise was planned and independently measured using the PTW Octavius(®) phantom and seven29(®) 2D array (PTW-Freiburg GmbH, Freiburg, Germany).
J BUON
November 2015
Department of Radiation Oncology, University of Texas Health Sciences Center at San Antonio, TX, USA.
Methods: Five patients with 38 fields have been analyzed in this study. The plans were optimized for the following clinical sites: one liver, one lung, one brain, one prostate and one spine. The detector array used for the measurements was the PTW Seven29 array.