Objective: This study compared the efficacy, safety, device satisfaction, and quality of life (QOL) in people with diabetes using an insulin bolus-patch versus current devices (pen/syringe) to deliver mealtime insulin.
Research Design And Methods: Thirty-eight subjects with diabetes (26 with type 1 and 12 with type 2) were randomized to bolus-patch or current injection device (55% pen and 45% syringe) to deliver mealtime insulin in a multicenter, 6-week crossover study. Efficacy was assessed by equivalence in mean daily seven-point blood glucose (MDBG). Safety assessments included severe hypoglycemia episodes, adverse device effects (ADEs), and adverse events (AEs). Device satisfaction was determined by the validated Insulin Delivery System Rating Questionnaire (IDSRQ) and QOL by the validated Diabetes Specific QOL Scale (DSQOLS).
Results: Using bolus-patch, MDBG (mean±SE) was equivalent to that using pen/syringe (8.61±0.28 vs. 9.02±0.26 mmol/L; P=0.098). SD of the seven-point blood glucose measurements was lower using bolus-patch (3.18±0.18 vs. 3.63±0.17 mmol/L; P=0.004), as was the coefficient of variation (CV) (37.2±1.7 vs. 40.3±1.7%; P=0.046). Hemoglobin A1c, 1,5-anhydroglucitol, fructosamine, and insulin use were similar between groups. There were no severe hypoglycemia episodes or serious ADEs. Between-device AEs were comparable. Subjects scored better on six of seven subscales on the DSQOLS and five of six subscales on the IDSRQ while using bolus-patch versus pen/syringe. At study completion, 76% of subjects would choose to switch to bolus-patch (P=0.001).
Conclusions: Delivery of mealtime insulin with bolus-patch compared with pen/syringe resulted in equivalent MDBG, lower SD and CV of seven-point blood glucose measurements, good safety, significant device satisfaction, and improved QOL.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4346544 | PMC |
http://dx.doi.org/10.1089/dia.2011.0047 | DOI Listing |
J Diabetes Sci Technol
July 2018
1 Profil, Neuss, Germany.
In the current issue of Journal of Diabetes Science and Technology, Dreon et al give a comprehensive overview of the technical performance of a wearable patch for bolus insulin delivery. The test results generated by the manufacturer of the bolus-patch provide the technical prerequisites for clinical application. As the device received FDA clearance already in 2010, positive results from the nonclinical performance testing were to be expected, but present nevertheless interesting insights into the device development.
View Article and Find Full Text PDFDiabetes Technol Ther
October 2011
Monteagle Medical Center, San Francisco, California 94110-4415, USA.
Objective: This study compared the efficacy, safety, device satisfaction, and quality of life (QOL) in people with diabetes using an insulin bolus-patch versus current devices (pen/syringe) to deliver mealtime insulin.
Research Design And Methods: Thirty-eight subjects with diabetes (26 with type 1 and 12 with type 2) were randomized to bolus-patch or current injection device (55% pen and 45% syringe) to deliver mealtime insulin in a multicenter, 6-week crossover study. Efficacy was assessed by equivalence in mean daily seven-point blood glucose (MDBG).