Purpose: End-stage renal disease patients with hemodialysis catheters in need of contrast enhanced imaging studies often have limited peripheral venous access. In this study we aimed to determine pressures generated in hemodialysis catheters during power injection of computed tomography (CT) contrast media.
Methods: Three different chronic hemodialysis catheters and two acute hemodialysis catheters were included in this study. All catheters were evaluated in vitro. A total volume of 120 cc of CT contrast material was injected at rate of 10 cc/s using a power injector. The catheters were connected to the power injector using a standard connecting tubing. Pressures were simultaneously measured in the power injector as well as in the hemodialysis catheters.
Results: The maximal measured pressures during injection in the power injector averaged 338 PSI (SD ± 8.7 PSI). The maximal measured pressure in the dialysis catheters ranged between 9.17 and 21.2 PSI. Pressures averaged 14.02 PSI (SD ± 3.34 PSI). The average pressure in the power injector was over 23 times higher than the pressure recorded at the hemodialysis catheter. None of the catheters ruptured or deformed during testing.
Conclusions: Pressures measured in hemodialysis catheters during power injection are lower than currently believed and markedly lower than the pressures recorded in the power injector. Standard hemodialysis catheters are likely to be amenable to power contrast injection in hemodialysis patients who have limited venous access. In vivo studies are necessary to confirm these findings.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.5301/JVA.2011.8442 | DOI Listing |
J Crit Care
January 2025
Division of Pulmonary and Critical Care Medicine, University of Maryland School of Medicine, Baltimore, MD, USA.
Purpose: This study evaluated the impact of choosing the right versus left internal jugular vein (IJV) for initial central venous catheter (CVC) placement on hemodialysis catheter-related outcomes in critically ill patients.
Materials And Methods: Medical records from the University of Maryland Medical Center were reviewed for adult critical care patients who received an IJV CVC between January 1, 2019, and December 31, 2022, and later required an additional temporary hemodialysis catheter.
Results: The study included 214 patients, with 100 (46.
Int J Artif Organs
January 2025
Konya Şehir Hastanesi, Konya, Turkey.
Background: Bleeding and thrombotic occlusion are complications of Central venous catheters. When selecting a catheter lock solution, factors such as bleeding, thrombotic occlusion, infection, and cost-effectiveness must be considered.
Methods: The study included 35 patients who used heparin as a locking solution and 35 patients who used 0.
BMC Nephrol
January 2025
Division of Nephrology, Department of Internal Medicine, Faculty of Health Sciences, University of the Free State, 205 Nelson Mandela Drive, Bloemfontein, 9300, South Africa.
Background: Continuous ambulatory peritoneal dialysis (CAPD) is one of the kidney replacement therapy (KRT) modalities used in patients with kidney failure. It is the preferred modality in most resource-limited settings as it is more accessible and cost-effective. CAPD technique failure remains a challenge and is associated with an increased risk of morbidity and mortality.
View Article and Find Full Text PDFNat Commun
January 2025
Querrey Simpson Institute for Bioelectronics, Northwestern University, Evanston, IL, USA.
Hemodialysis for chronic kidney disease (CKD) relies on vascular access (VA) devices, such as arteriovenous fistulas (AVF), grafts (AVG), or catheters, to maintain blood flow. Nonetheless, unpredictable progressive vascular stenosis due to neointimal formation or complete occlusion from acute thrombosis remains the primary cause of mature VA failure. Despite emergent surgical intervention efforts, the lack of a reliable early detection tool significantly reduces patient outcomes and survival rates.
View Article and Find Full Text PDFIntroduction: The lack of peritoneal dialysis (PD) catheters designed explicitly for neonates creates significant challenges in the provision of neonatal peritoneal dialysis. High resource settings can circumvent this limitation by resorting to alternative extracorporeal dialysis methods. However, low resource settings (LRS), PD remains the preferred dialysis modality, and the use of off-label catheters for PD results in complications such as omental wrapping and occlusion.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!