The aim of this study was to investigate the dosimetric characteristics of the electron beams generated by the light intraoperative accelerator, Liac® (SORDINA, Italy), using Monte Carlo (MC) calculations. Moreover we investigated the possibility of characterizing the Liac® dosimetry with a minimal set of dosimetric data. In fact accelerator commissioning requires measurements of both percentage depth doses (PDDs) and off-axis profiles for all the possible combinations of energy, applicator diameter and bevelled angle. The Liac® geometry and water phantom were simulated in a typical measurement setup, using the MC code EGSnrc/BEAMnrc. A simulated annealing optimization algorithm was used in order to find the optimal non-monoenergetic spectrum of the initial electron beam that minimizes the differences between calculated and measured PDDs. We have concluded that, for each investigated nominal energy beam, only the PDDs of applicators with diameters of 30, 70 and 100 mm and the PDD without an applicator were needed to find the optimal spectra. Finally, the output factors of the entire set of applicator diameters/bevelled angles were calculated. The differences between calculated and experimental output factors were better than 2%, with the exception of the smallest applicator which gave differences between 3% and 4% for all energies. The code turned out to be useful for checking the experimental data from various Liac® beams and will be the basis for developing a tool based on MC simulation to support the medical physicist in the commissioning phase.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0031-9155/56/14/022DOI Listing

Publication Analysis

Top Keywords

monte carlo
8
electron beams
8
beams generated
8
find optimal
8
differences calculated
8
output factors
8
carlo simulation
4
simulation electron
4
generated mev
4
mev dedicated
4

Similar Publications

To estimate the cost-effectiveness of cetuximab in combination with radiotherapy compared with radiotherapy alone, for the treatment of locally advanced head and neck cancer patients in Spain. A probabilistic Markov model (second-order Monte Carlo simulation) with a five-year time horizon and quarterly Markov cycles was performed from the perspective of the Spanish National Health System (NHS). The additional cost and quality-adjusted life-year (QALY) gain per patient receiving radiotherapy in combination with cetuximab compared with radiotherapy alone was €4356 (95% CI: €4350-4362) and 0.

View Article and Find Full Text PDF

Assessment of the Performance of the Dose Calibrator Used in Radioactivity Measurement.

Indian J Nucl Med

November 2024

Center for Research and Production of Radioisotopes, Dalat Nuclear Research Institute, Vietnam Atomic Energy Institute (VINATOM), Da Lat City, Lam Dong Province, Vietnam.

Aims: This study aimed to evaluate the principal technical characteristics of a well-type gas-filled ionization chamber dose calibrator used in measuring radiopharmaceutical activity, namely accuracy, repeatability, and linearity. Furthermore, this work also explored the correlation between the device's response and the position and volume of the radiopharmaceutical I-131.

Materials And Methods: Experimental measurements were conducted on the ATOMLAB 500 dose calibrator using NIST traceable Cs-137 source to determine the accuracy and repeatability.

View Article and Find Full Text PDF

The abandoned coal in goaf will adsorb the gases ethylene (CH) and acetylene (CH) produced by coal oxidation, which makes the concentration data of the indicator gas inaccurate. Therefore, the adsorption law of coal and CH and CH gas is explored. The macromolecular structure model of coal was established and optimized by simulation, and the gas adsorption process was simulated by means of grand canonical Monte Carlo method.

View Article and Find Full Text PDF

Comparative Study of Polymer Globules and Liquid Droplets in Poor Solvents: Effects of Cosolvents and Solvent Quality.

J Phys Chem B

January 2025

Department of Chemical Engineering, IIT Gandhinagar, Gandhinagar, Gujarat 382055, India.

We compare the structures of polymer globules, composed of flexible polymer chains, with liquid droplets made of nonbonded monomers of the same polymer in poor solvents. This comparison is performed in three different poor solvents, with and without the addition of cosolvents. Molecular dynamics simulations are used to analyze the properties of the polymer globules, while semigrand canonical Monte Carlo simulations are used to form metastable liquid droplets of nonbonded monomers through homogeneous nucleation in the same solvents.

View Article and Find Full Text PDF

Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.

Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!