Background: Stress hyperglycemia in the critically ill has been found to be associated with increased morbidity and mortality. Studies have found significant improvements in morbidity and mortality in postsurgical patients whose glucose levels were closely maintained in the euglycemic range. However, subsequent studies, in particular the Normoglycemia in Intensive Care Evaluation and Survival Using Glucose Algorithm Regulation (NICE-SUGAR) study, found no improvement in subjects with tight glycemic control. In addition to differences in protocol design, patients in the tight glycemic control arm of the NICE-SUGAR study experienced high rates of hypoglycemia compared with other studies. One interpretation of the NICE-SUGAR study results is that it is difficult to achieve normal glycemia in critically ill patients with existing glucose monitoring technology. The purpose of the study reported here was to evaluate the safety and performance of a continuous intravascular glucose sensor that could be used in the future in critically ill patients.
Methods: A first-generation prototype of an intravascular continuous glucose sensor was evaluated in 29 volunteer subjects with type 1 diabetes mellitus. The sensor operates on the principle of quenched fluorescence. The fluorescent emission from the sensor chemistry is nonlinear, resulting in improved accuracy in the hypoglycemic range. The duration of each study was 8 hours. Sensor output was compared with temporally correlated reference measurements made from venous samples on a laboratory glucose analyzer.
Results: Data were obtained from 18 of the 29 subjects in the study. Data were analyzed retrospectively using a factory calibration plus a one-point in vivo calibration. The mean absolute relative difference was 7.97%, and 95.1% of all the points were in zone A of the Clarke error grid.
Conclusions: This pilot study was the first use in human subjects of a prototype of the GluCath Intravascular Continuous Glucose Monitoring System (GluCath System). The GluCath System is based on a novel fluorescent sensor chemistry. The study found the GluCath System had a high level of accuracy as compared with a laboratory reference analyzer.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3192635 | PMC |
http://dx.doi.org/10.1177/193229681100500323 | DOI Listing |
Anal Chim Acta
May 2025
College of Textile Science and Engineering, Jiangnan University, 1800Lihu Road, Wuxi, 214122, China. Electronic address:
Flexible sweat sensors possess the special potential for continuous non-invasive monitoring of human blood glucose. We put forward a flexible microcolumn array sensor, which is designed for health monitoring by means of detecting glucose levels in sweat and capturing physiological signals related to human movement. With the combination of silk fibroin (SF), waterborne polyurethane (PU), and multi-walled carbon nanotubes (MWCNT), this microcolumn film electrode is able to effectively function as a strain sensor benefiting from the superior mechanical performance of PU.
View Article and Find Full Text PDFBiol Pharm Bull
March 2025
Department of Pharmacy, College of Pharmacy, Kinjo Gakuin University, 2-1723 Omori, Moriyama-ku, Nagoya 463-8521, Japan.
We examined whether the glucose levels and awareness of individuals without diabetes changed after using a sensor-based intermittently scanned continuous glucose monitoring (isCGM) system in their daily lives. Japanese individuals without a diabetes diagnosis wore the isCGM system while maintaining a normal lifestyle during the baseline period. A certified diabetes educator coached them on how to improve their lifestyle based on information from sensor data, food journals, and body composition.
View Article and Find Full Text PDFJ Clin Endocrinol Metab
March 2025
Department of Metabolic Medicine, Faculty of Life Sciences, Kumamoto University. 1-1-1 Honjo, Chuo-ku, Kumamoto 860-8556, Japan.
Context: In 2023, we employed Dexcom G6 for real-time continuous glucose monitoring (rt-CGM); it showed high usefulness but unsatisfactory accuracy in type 1 diabetes summer camp (camp) participants.
Objective: To assess the usefulness, recommendation and accuracy of a new rt-CGM system in camp, 2024.
Methods: Sensor glucose (SG) concentrations were measured by Dexcom G7 (G7) from 6 days prior to camp.
Mikrochim Acta
March 2025
School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, 20093, China.
A disposable, self-powered enzymatic biofuel cell (BFC) sensor integrated with a hollow microneedle array (HMNA) for glucose monitoring in interstitial fluid (ISF) is reported. The HMNA enables painless and minimally invasive ISF extraction. The BFC uses dehydrogenase (GDH) in conjunction with NAD, diaphorase (DI), and vitamin K (VK) serving as electron transfer mediators as the anode catalyst and Prussian blue (PB) as the electrochromic cathode.
View Article and Find Full Text PDFPhysiol Rep
March 2025
Renal Research Institute, New York City, New York, USA.
Monitoring intraperitoneal pressure (IPP) offers valuable insights into changes of intraperitoneal volume (IPV) during peritoneal dialysis (PD). This study aims to investigate the relationship between IPV and IPP during a PD dwell. Thirteen patients were studied during a 2-h dwell using 2 L of dialysate containing 2.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!