Pound-Drever-Hall-locked, frequency-stabilized cavity ring-down spectrometer.

Rev Sci Instrum

Instytut Fizyki, Uniwersytet Mikołaja Kopernika, ul. Grudziadzka 5/7, 87-100 Toruń, Poland.

Published: June 2011

We describe a high sensitivity and high spectral resolution laser absorption spectrometer based upon the frequency-stabilized cavity ring-down spectroscopy (FS-CRDS) technique. We used the Pound-Drever-Hall (PDH) method to lock the probe laser to the high-finesse ring-down cavity. We show that the concomitant narrowing of the probe laser line width leads to dramatically increased ring-down event acquisition rates (up to 14.3 kHz), improved spectrum signal-to-noise ratios for weak O(2) absorption spectra at λ = 687 nm and substantial increase in spectrum acquisition rates compared to implementations of FS-CRDS that do not incorporate high-bandwidth locking techniques. The minimum detectable absorption coefficient and the noise-equivalent absorption coefficient for the spectrometer are about 2×10(-10) cm(-1) and 7.5×10(-11) cm(-1)Hz(-1/2), respectively.